Vante
™
Model 3120
9
3.
Maintenance and Repair
3.1
System Maintenance
3.1.1
The only maintenance required for the model 3120 RF Generator is
keeping the fan intake filter and air exhaust ports free of accumulated dust to
ensure proper airflow.
WARNING
: Turn off the Generator and disconnect the sealer by removing all
connectors prior to cleaning. Failure to do so may result in an RF
burn to the operator during the cleaning process.
3.2
Repair
3.2.1
The replacement of a fuse, located in the AC input receptacle on the back
of the unit, is a rare occurrence. If the unit is plugged into an AC power source
and the AC power-on light is not illuminated when the rocker switch is in the on
position, then one of the fuses
may
need replacement.
3.2.2
To access the fuse compartment the AC cord must be unplugged. Release
the fuse holder slide latch by inserting the end of a flat blade screwdriver into the
detent and sliding the fuse holder out until the fuses are exposed (Figure 1.2).
3.2.3
Examine the fuses and replace as required. If the fuses are not blown,
check again to be sure all other connections are correct.
Be sure to replace fuses
with one of the specified rating!
If fuses continue to blow, contact Vante™.
3.2.4
For all other questions and/or problems, call Vante™ directly at 520-881-
6555 or 877-565-5557.
NOTE:
Units returned to Vante™ for repair are subject to biohazard charges if
any Selaer component is contaminated with media or other potentially
biohazardous materials.
Содержание 3120
Страница 1: ...Doc No 31200801 00 Rev A INSTRUCTION MANUAL MODEL 3120 200 Watt RF Generator...
Страница 2: ...ii This page intentionally left blank...
Страница 4: ...iv This page intentionally left blank...
Страница 13: ...Vante Model 3120 5 Figure 1 3 Serial Plate Figure 1 4 Fuse Label...
Страница 14: ...Vante Model 3120 6 This page intentionally left blank...
Страница 18: ...Vante Model 3120 10 This page intentionally left blank...