Vanguard Instruments VBT-75P S2 Скачать руководство пользователя страница 19

VBT-75P S2 USER’S MANUAL  REV 1.0 

 

 

 

 

16 

 

j.

 

The test results will be printed, and the following screen will be displayed: 

 

Press the 

[1]

 key (

YES

) to keep the test reading. 

k.

 

The following screen will be displayed: 

 

Press any key to continue. 

l.

 

The following screen will be displayed: 

 

Press the 

[2]

 key (

NO

). 

m.

 

The following screen will be displayed: 

 

Press the 

[1]

 key (

YES

) to save the test record. 

n.

 

The following screen will be displayed: 

 

Press any key to return to the "START-UP" menu. 

 

RECORD 6   SAVED! 
 
 
 

SAVE THIS RECORD? 
1.YES 
2.NO 

 

RUN ANOTHER TEST? 
1.YES 
2.NO 
 

 

TEST SAVED 

 
 

KEEP THIS READING? 
1.YES 
2.NO 

 

Содержание VBT-75P S2

Страница 1: ...VBT 75P S2 VACUUM BOTTLE TESTER USER S MANUAL Vanguard Instruments Company Inc 1520 S Hellman Ave Ontario California 91761 USA TEL 909 923 9390 FAX 909 923 9391 July 2017 Revision 1 0...

Страница 2: ...el or by an authorized repair service provider Unauthorized modifications can cause safety hazards and will void the manufacturer s warranty EQUIPMENT RATINGS IP Rating The enclosure for the VBT 75P S...

Страница 3: ...the VBT 75P S2 s Memory 18 3 8 Printing a Directory of Test Records Stored in a USB Flash Drive 20 3 9 Reviewing a Test Record 21 3 10 Saving a Test Record 23 3 11 Restoring a Test Record from the Uni...

Страница 4: ...o SPLAY TE SPLAY TE SPLAY TE SPLAY TE S2 USER S M CONVENT the followin on the VBT 7 u names are screen outpu es are indica arning mess are indicate te details XT LINE XT LINE XT LINE XT LINE MANUAL TI...

Страница 5: ...o illuminated The presence of high voltage is indicated by an audible tone and an illuminated HIGH VOLTAGE LED light For additional operator safety an ARM switch must be held down during testing The V...

Страница 6: ...d paper jams the use of thermal paper supplied by Vanguard Instruments Company is highly recommended Additional paper can be ordered from the following sources Vanguard Instruments Co Inc 1520 S Hellm...

Страница 7: ...cations are 25 C 77 F S2 Technical vacuum bottle 7 W x 10 5 H x 5 44 Kg 2A 50 60 Hz dc in 1 kV steps harge time for i 0 characters x tor LED illumi A programm ndicator LED membrane key records of 16 d...

Страница 8: ...functional description in Table 2 The table describes the function of each item on the control panel The purpose of the controls and indicators may seem obvious but users should become familiar with...

Страница 9: ...and power switch 5 PUSH TO ARM Arm switch press and hold during testing 6 TEST FAILED Test failure indicator This indicator turns on if the test current exceeds the present current threshold 100 200 o...

Страница 10: ...AL TIONS d with one 1 s are termin ed A typical BT 75P S2 ag s ground stu er must be o e under test re 2 VBT 75P 0 foot 3 05 ated with al cable conne ainst static d ud to the sub off line and c t shou...

Страница 11: ...will be displayed Enter the date and time using the membrane keypad The date and time will be set and you will be returned to the START UP menu 3 2 Adjusting LCD Screen Contrast To adjust the LCD scre...

Страница 12: ...mpany name using the keypad When pressing a key the corresponding number on the key will be displayed first Pressing the key again will display the first letter on the key Pressing the key again will...

Страница 13: ...ENTER key g The following screen will be displayed Type the model information using the keypad and then press the ENTER key h The following screen will be displayed Type the serial number using the ke...

Страница 14: ...ANUAL 11 j The following screen will be displayed Type the operator s name using the keypad and then press the ENTER key All header information will be saved and you will be returned to the START UP m...

Страница 15: ...layed Press the 3 key SET LANGUAGE c The following screen will be displayed Select the preferred language by pressing the corresponding number on the keypad d A confirmation message will be displayed...

Страница 16: ...CAL PRINT ON OFF c The following screen will be displayed 1 Press the 1 key to enable the option The following screen will be displayed The calibration date printing option will be enabled and you wi...

Страница 17: ...st voltage using the numeric keypad and then press the ENTER key d The following screen will be displayed Select the leakage current limit by pressing the corresponding number on the keypad For this e...

Страница 18: ...st voltage is returned to zero and a PASS message is displayed as shown below However if a flash over condition occurred such as bottle failure the test voltage is immediately turned off and a FAIL me...

Страница 19: ...ll be displayed Press any key to continue l The following screen will be displayed Press the 2 key NO m The following screen will be displayed Press the 1 key YES to save the test record n The followi...

Страница 20: ...REV 1 0 17 N VBT 75P S OTE The S2 USER S M Fig e VBT 75P S2 MANUAL ure 3 Samp 2 can store u le Test Resu up to 84 rec ults Printout ords of 16 re eadings each h...

Страница 21: ...OR L DIRECTO 1 INTERNA 2 THUMB D OPERATIO 2 REST W 4 DIR EST 07 1 13 3 DS VTG 0 of Test Rec he test recor P menu RDS be displaye be displaye drive is inser en will be dis ey INTERNA print the test y l...

Страница 22: ...ay the next or previous test record respectively The test record information will be displayed as shown Press the STOP key to return to the START UP menu Figure 4 Sample Test Record Directory Printout...

Страница 23: ...tart from the START UP menu Press the 3 key RECORDS c The following screen will be displayed Press the 4 key DIR d The following screen will be displayed Press the 2 key THUMB DRIVE DIR e The Flash dr...

Страница 24: ...will be displayed Press the 1 key SCROLL TEST RECORD if you would like to review the test record in the working memory on the unit s LCD screen Continue to step d Press the 2 key PRINT TEST RECORD to...

Страница 25: ...he following screen will be displayed Press any key to continue f The following screen will be displayed Press the STOP key to return to the START UP menu TEST NUMBER 1 75KV 5Sec 300 A 185 A AT 74 9KV...

Страница 26: ...START UP menu Press the 3 key RECORDS b The following screen will be displayed Press the 1 key SAVE If a USB Flash drive is connected to the unit continue to step c If a USB Flash drive is NOT connec...

Страница 27: ...al memory The following screen will be displayed Press any key to return to the START UP menu 2 SAVE TO THUMB DRV Press the 2 key SAVE TO THUMB DRV to save the data in the unit s working memory to a t...

Страница 28: ...cor ECORD NUM record num screen will STORE REC UMBER ORE RECO RECORD N TO SELE 1 INTERNA 2 THUMB D OPERATIO 2 REST W 4 DIR EST 07 1 13 3 DS VTG 0 MANUAL ord from th re a test reco P menu RDS be displa...

Страница 29: ...e displayed Press either the key or the key to display the next or previous test record respectively The test record information will be displayed as shown When you have located the test record that y...

Страница 30: ...splayed Press the 2 key RESTORE d The following screen will be displayed Press the 2 key THUMB DRIVE e The following screen will be displayed Type the test record number you would like to restore and...

Страница 31: ...e displaye choose to pr he unit s wor n 3 9 S to print th al printer a O if you do n START UP m D d int or not pr rking memor he restored t nd you will b not want to menu VBT 75P S2 rint the resto ry...

Страница 32: ...splayed 1 COPY SINGLE RECORD Press the 1 key COPY SINGLE RECORD if you would like to copy a single test record to the USB Flash drive The following screen will be displayed Type the test record number...

Страница 33: ...ALL RECORDS Press the 2 key COPY ALL RECORDS to copy all test records from the unit s internal memory to the connected USB Flash drive The test records will be copied and you will be returned to the...

Страница 34: ...lowing scree ess the 1 k NGLE REC 1 key if yo yed record num ASE RECO MBER SE RECOR SINGLE R ALL RECO 1 INTERNA 2 THUMB D OPERATIO 2 REST W 4 DIR EST 07 1 13 3 DS VTG 0 MANUAL s from the e one or all...

Страница 35: ...ld lik M The follo ANCEL the e u will be ret key to proce le the record will be displ rn to the ST RASED SHOTS CORDS URE NTINUE RASED STOP key t the record n nstructions layed TART UP m e to erase a o...

Страница 36: ...layed Press the 5 key ERASE d The following screen will be displayed Press the 2 key THUMB DRIVE e The following screen will be displayed 1 ERASE SINGLE REC Press the 1 key if you would like to erase...

Страница 37: ...ould lik drive The fo ANCEL the e u will be ret key to proce le the record will be displ rn to the ST RASED SHOTS CORDS URE NTINUE REC_007 STOP key t the record n nstructions layed TART UP m e to eras...

Страница 38: ...will need to be approved first before it can be used to download any software or firmware This is usually done within 30 minutes of creating your account b After logging in to your account hover your...

Страница 39: ...m fr ommend us ned off inse STOP key hrough the in ss The STA root folder o ty flash drive rom the driv ing a flash d ert the USB F y turn on th nitial startin RT UP men VBT 75P S2 of a USB Flas e Ple...

Страница 40: ...S Hellman Ave Ontario CA 91761 USA Phone 909 923 9390 Fax 909 923 9391 www vanguard instruments com Copyright 2017 by Vanguard Instruments Company Inc VBT 75P S2 User s Manual Revision 1 0 July 27 201...

Отзывы: