background image

     

ATO-

250™ Operating Procedures 

 

 

10.0 Operating Procedures 

Before novice operators use the ATO-250 to 
measure any resistance, they should review 
Figure 5.0, the 

Step-by-Step Operating 

Procedures 

for

 

the ATO-250.  The figure 

shows the flowchart of the branching of the 
multiple test options.  The ATO-250 
operation is simple, thus requiring minimal 
input. The operator only needs to select 
choices from the display menus and 
responding to the displayed prompts. 
 

10.1 Precautions 

The purpose of the ATO-250 is to measure 
non-inductive low resistance devices such as 
high-power, circuit-breaker contacts. 

 

WARNING 

Do  not  measure  resistance  of  any 
inductive  device,  which  will  create  an 
unsafe,  high-voltage  spike  due  to  the 
collapsing  of  the  magnetic  field  if  the 
test  current  is  interrupted  by  a 
detachment  of  a  test  lead.  Do  not 
disconnect  or  touch  any  test  lead 
connected  to  a  device  being  tested 
while  current  is  being  conducted.  

Failure  to  heed  this  warning  can 
result  in  personal  injury  and  damage 
to the ATO-250. 

If the resistance of an inductive device 
is desired, Vanguard Instruments does 
manufacture the WRM, an instrument 
designed for that purpose.  

 

10.2 Preparations 
a. 

Plug the ATO-250 power cable into a 

120-volt power outlet. If the unit is setup for 
240-volt operation, plug the power cable 
into a 240-volt outlet receptacle. 
 

b. 

Insert the test-current cable plugs and the 

test voltage-sensing cable plugs into the 
respective control-panel jacks. 
 

c. 

Attach the opposite terminals of the test-

cable clamps to the resistive load that is to 
be tested. (See Figure 4.0) 
 

NOTE 

It is important that the test clips be securely 
attached to the contact bushings, bus bar or 
junction terminals to eliminate errors from 
contact resistance, due to the possibility that 
the test currents can be as high as 250 
Amperes. 
 

d. 

To turn on the ATO-250’s power, press 

I

” on the rocker switch. (Item 4 in Figure 

1.0) 
Observe that after the configuration data is 
displayed briefly, then the START UP menu 
is shown: 

 

 
 
 
 
 

 

11.0 Run Test Procedure 

A resistance may be measured by 
performing the following step-by-step 
sequence. However, if a test is to be fully 
documented, the identification data must 
also be entered for the equipment that is to 
be tested. See Section 12.0 for entering the 
identification data. Since the test time and 
date will be stored with test record, the clock 
and date must be also be correct before any 
test is performed. 
 

a. 

Press the number one key on the START-

UP menu. The CONNECT CABLES prompt 
will display: 
 
 
 
 
 
 
 

b. 

Slowly turn the CURRENT CONTROL 

knob until the desired test current is shown. 

CONNECT CABLES THEN 

ADJUST DRIVE TO 

DESIRED CURRENT 

(250 A MAX)

 

1.RUN TEST           05/15/01 
2.SETUP                 10:28:23 
3.CAL CHECK 

Содержание II Series

Страница 1: ...S for the MODEL ATO 250 10 250 Amperes Digital Micro Ohm Meter Series II ID Number VIC 25000 S2 Vanguard Instruments Company 1710 Grevillea Court Ontario California 91761 TEL 909 923 9390 JULY 2001 FA...

Страница 2: ...isk of introducing additional or unknown hazards To ensure that all designed safety features are maintained it is recommended that repairs be performed only by Vanguard Instruments Company factory per...

Страница 3: ...s 2 6 0 Controls and Indicators 3 7 0 Preset Setup 5 8 0 ATO 250 Printer Paper 6 9 0 ATO 250 Cable Connections 6 10 0 Operating Procedures 8 11 0 Run Test Procedure 8 12 0 Enter Test Information Proce...

Страница 4: ...5 Figure 4 0 ATO 250 Cable Connection 6 Figure 5 0 Step by Step Operating Procedures 7 Figure 6 0 Typical ATO 250 Test Report 11 Figure 7 0 Typical Test Record Directory Print Out 15 Figure 8 0 Bus Ba...

Страница 5: ...ows based software package is supplied with each ATO 250 that allows the user to retrieve test records stored in the unit s memory Operation requires little more than connecting the test leads to an u...

Страница 6: ...hm meter POWER 85 to 132 Vac or 200 240 Vac selectable 50 60 Hz SIZE inches 16 8 wide by 12 6 high by 10 6 deep WEIGHT Less than 42 lbs TEST CURRENT 10 to 250 amperes READING RANGE 1 micro ohm to 300...

Страница 7: ...mber is cross referenced to a functional description in Table 2 0 Although the purpose of these controls and the display may seem obvious and intuitive users should become familiar with them before at...

Страница 8: ...el from 10 to 250 Amperes 7 None 4 line by 20 character backlit LCD screen Use for displaying menu selections prompts and test result data 8 None Keypad consisting of momentary contact pushbuttons Ten...

Страница 9: ...ction is set by jumpers on a terminal block as shown in Table 3 0 and in Figures 2 0 and 3 0 Table 3 0 Input Power Voltage Selections VOLTAGE SELECTION TERMINAL BLOCK JUMPERS 90 130 Vac Brown to Blue...

Страница 10: ...nt Co 13607 E Trent Avenue Spokane WA 99216 Tel 509 893 9881 Fax 509 893 9803 Part Number TP 3 paper 9 0 ATO 250 Cable Connections The ATO 250 is supplied with a set of 35 foot test cable Each cable c...

Страница 11: ...ATO 250 Operating Procedures 7 Figure 5 0 Step by Step Operating Procedures...

Страница 12: ...setup for 240 volt operation plug the power cable into a 240 volt outlet receptacle b Insert the test current cable plugs and the test voltage sensing cable plugs into the respective control panel ja...

Страница 13: ...Figure 6 After the test result is printed the KEEP THIS READING prompt will appear If a printout of the test record is not needed then press the number two key NO in which case the following prompt i...

Страница 14: ...TEST then the SAVE THIS RECORD display is shown next o To save the record into the FLASH EEPROM select the number one key YES The ATO 250 will assign a record number automatically as shown below p Th...

Страница 15: ...key once to select the number 2 Press the key a second time to select the letter A If the key is pressed a third time letter B will be selected While the fourth time will select the letter C Pressing...

Страница 16: ...ers for the SERIAL NUMBER Then press the ENTER key The KVA RATING prompt will display i Press the characters of the KVA RATING Then press the ENTER key The OPERATOR ID prompt appears next j Place the...

Страница 17: ...ich will printout the test record and return to the START UP menu h When the list scrolls to the record of interest shown on the above display press the ENTER to display the date and time of the test...

Страница 18: ...ER CONTROL to allow the ATO 250 to be controlled by an IBM compatible personal computer The computer interface ITF mode advisory will display d To end the Computer Interface mode press the STOP key at...

Страница 19: ...printout the display returns to the START UP menu A typical test record directory print out is shown in Figure 7 0 Figure 7 0 Typical Test Record Directory Print Out 16 0 Erase Record Procedure The fo...

Страница 20: ...allows the user to verify the electronics of the ATO 250 is working properly All of the procedures start at the START UP menu shown below a Press the number three key to begin the CAL CHECK procedure...

Страница 21: ...s the following test results will display in rapid succession f Press the ENTER key to return the display to the START UP menu This completes the CAL CHECK procedure Figure 8 0 Bus Bar Connection ZERO...

Страница 22: ...urrent teeth are touching the device under test DUT 2 Inspect sensing cables 2 No Test Current Current reads zero and resistance reading is also zero 1 No test current going through the DUT 2 Drive ci...

Страница 23: ...Operating Procedures 19 APPENDIX B ATO 250 Cable Accessories Plastic Quick Disconnect Clip C Clamp Current Connector and Sensing Wire Connector C Clamp Plastic Clip Current Connector Sensing wire Con...

Страница 24: ...ATO 250 Operating Procedures 1710 Grevillea Court Ontario CA 91761 USA Phone 909 923 9390 Fax 909 923 9391 Website http www vanguard instruments com ATO 250 11 05 01 OAK...

Отзывы: