![Vanguard Instruments EZCT S2 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/vanguard-instruments/ezct-s2/ezct-s2_user-manual_3388302021.webp)
REV 3
EZCT S2 USER’S MANUAL
17
NOTE
The “HIGH VOLTAGE PRESENT” light will be illuminated to indicate that high
voltage is present.
k.
If the selected test included a ratio test, the following screen will be displayed:
When the ratio test is complete, the final results will be displayed:
Press any key on the keypad to continue.
l.
The following screen will be displayed:
Press the
[1]
key (
YES
) if you would like to print the test results. The test results will be
printed on the thermal printer. A typical EZCT S2 tabulated test report printout is shown
in Figure 5. A typical graphic report is shown in Figure 6.
Press the
[2]
key (
NO
) if you do not want to print the test results.
m.
The following screen will be displayed:
Press the
[1]
key (
YES
) to keep the test results.
KEEP THIS TEST?
1. YES
2. NO
PRINT TEST RESULTS?
1. YES
2. NO
ANY KEY TO CONTINUE
Vx= 39.9 Vh= 0.4984
I=0.1324 RAT=+80.058
RATIO TEST
Vx= 39.9 Vh= 0.4984
I=0.1324 RAT=+80.058
Содержание EZCT S2
Страница 23: ...REV 3 EZCT S2 USER S MANUAL 19 Figure 5 Typical EZCT S2 Tabulated Report Printout...
Страница 25: ...REV 3 EZCT S2 USER S MANUAL 21 Figure 6 Typical EZCT S2 Graphic Report Knee Point Marker...
Страница 31: ...REV 3 EZCT S2 USER S MANUAL 27 Figure 7 Typical Internal Test Record Directory Printout...