Response:
STX
15
SN
XXXXXXXXXX
CC
Data field length is 10 characters
XXXXXXXXXX = Serial number 10 characters
5.4.4
Read Real Time Pressure and Temperature Value
Read a real-time pressure and temperature value of the Diver including a time stamp. If this command
is given the Diver will take a reading immediately whether the Diver is logging or not. This data will not
be stored in the Diver memory.
Command:
STX
5
RT
none
CC
Response:
STX
42
RT
YY/MM/DD HH:MM:SS:XXXXX.XXX:ZZZZZ.ZZZ
CC
Data field length is 37 characters per described format
YY/MM/DD HH:MM:SS = Date/time format 17 characters
XXXXXX.XXX = value Level (in cmH
2
O with 3 decimals) 9 characters
ZZZZZZZ.ZZZ = value Temperature (in degrees Celsius with 3 decimals) 9 characters
5.4.5
Read Recorded Pressure and Temperature
Read the data recorded by the Diver. Each data record consisting of a time stamp, pressure and
temperature value must be read separately.
Command:
STX
15
SD
XXXXXXXXXX
CC
Data field length is 10 characters per described format
XXXXXXXXXX = Record number 10 characters (first record is record number 1; last record number is
72,000)
Example
Description
BX270
␣␣␣␣␣
10 characters (all ASCII)
Example
Description
22/05/06
␣
17:25:26:
␣
1445.166:
␣␣␣
22.273
37 characters (all ASCII)
Example
Description
10000
␣␣␣␣␣
10 characters (all ASCII)