Pressure
The specifications for atmospheric and water pressure measurements vary by type of Diver. The
specifications below apply at operating temperature.
TD-Diver
DI801
DI802
DI805
DI810
Unit
Water column measurement range
10
20
50
100
mH
2
O
Accuracy (max)
± 2.0
± 4.0
± 10.0
± 20.0
cmH
2
O
Accuracy (typical)
± 0.5
± 1.0
± 2.5
± 5.0
cmH
2
O
Long-term stability
± 2
± 4
± 10
± 20
cmH
2
O
Resolution
0.2
0.4
1
2
cmH
2
O
Display resolution
0.058
0.092
0.192
0.358
cmH
2
O
Overload pressure
15
30
75
150
mH
2
O
Baro-Diver
DI800
Unit
Water column measurement range
1.5
mH
2
O
Accuracy (max)
± 2.0
cmH
2
O
Accuracy (typical)
± 0.5
cmH
2
O
Long-term stability
± 2
cmH
2
O
Resolution
0.1
cmH
2
O
Display resolution
0.058
cmH
2
O
Overload pressure
15
mH
2
O
2.5.1
Water Column Measurement Range
The height of water above the Diver that can be measured.
2.5.2
Accuracy (maximum)
Accuracy is the proximity of measurement results to the true value. The algebraic sum of all the errors
that influence the pressure measurement. These errors are due to linearity, hysteresis and
repeatability. During the Diver calibration process a Diver is rejected if the difference between the
measured pressure and the applied pressure is larger than the stated accuracy.
2.5.3
Accuracy (typical)
At least 68% of the measurements during the calibration check are within 0.05% FS of the
measurement range.