SET-UP 
 
CONNECTING PARTS OF THE SpinVessel® SYSTEM 
 
Power and Controller Connections 
 

 

 

 

 

1.

 

Place Motor Unit on a sturdy lab bench, table, or robotic workstation.  Note that Z height 

with tube is 136mm (rubber feet not included). 

2.

 

Since Controller Cable is 6 feet long, Controller and computer can be placed at a distance 

from the Motor Unit.  

3.

 

Connect Power Cord from Power Supply to an outlet. 

Power Switch 

Power Cable 

Connetor 

USB 

Cable Connector 

Controller 

Cable 

Connector 

Controller 

Controller 

Cable 

Controller 

Motor 

Unit 

Power 

Supply 

Power 

Cable 

Power Cord 

USB 

Cable

 

Spin 

Base 

USB 

Drive

 

Содержание SpinVessel VP 418SV2-1L-FB-CC

Страница 1: ...d Foreign Patents Pending SpinVessel System with Computer Control feature for 1L SpinVessel Tube VP 418SV2 1 1L FB CC includes Motor Unit Spin Base Controller Power Supply not shown Registration Disc...

Страница 2: ...ght with tube is 136mm rubber feet not included 2 Since Controller Cable is 6 feet long Controller and computer can be placed at a distance from the Motor Unit 3 Connect Power Cord from Power Supply t...

Страница 3: ...troller plugged in before powering up 5 Connect Controller Cable from the Motor Unit to the Controller 6 Connect the USB Cable from the Controller to the computer 7 Place one VP 830SV 650CFB SpinVesse...

Страница 4: ...nVessel tube in the drive base check the tip alignment relative to the projections Without liquid in the tube turn on the SpinVessel to make sure the tips are not hitting the projections Insert the Sp...

Страница 5: ...r software installed If not download the CH340 USB DRIVER from the provided USB flash drive and run CH341SER exe to install it Windows typically auto installs once downloaded If your company policy do...

Страница 6: ...Create ASCII command strings as outlined in the diagram shown below to control the instrument with the following parameters a Speed of rotation in revolutions per minute RPM b Number of rotations bef...

Страница 7: ...enomenon The following settings are ones to use when first testing the VP 418SV 1 1L FB CC SpinVessel Speed 150 RPM Rotate 2 1 Pause 200 milliseconds Ramp 10 It is always a good idea to practice with...

Страница 8: ...tor control product Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design manufacture and intended use of the products V P Sci...

Страница 9: ...ow for return authorization and shipping instructions This warranty is made in lieu of other warranties expressed or implied including the warranties of merchantability and fitness for a particular pu...

Отзывы: