Compact Turbimeter

11

VAN WALT

Monitoring your needs

 

Readings are consistently higher than expected :

Check the sample cuvette for dirt, scratches and marks. Turbidity is an optical 

measurement and will be adversely affected by unwanted items in the optical 

paths.

Index the cuvette using the Continuous-Capture mode to ensure the most accurate 

reading position is selected.

Apply a thin film of silicone oil to the cuvette to ‘fill in’ any scratches.

Replace the cuvette with a new, unused cuvette.

Check the calibration using the supplied standards to ensure no offset has been 

inadvertently applied.

 

Readings are consistently lower than expected :

Fouling in the optical chamber may prevent the sources being transmitted with full 

intensity. The unique QuadoptiX™ system will prevent this issue arising in all but the 

most challenging circumstances.

Gently clean the optical chamber using a  oft, damp cloth. Do not use abrasive 

materials.

Readings are unstable/variable :

Turbidity is a measurement of scattered light with the source of scattering being 

microscopic and larger solids present in the sample. Solids can be mobile or settle 

during measurement leading to changes in readings over time. 

Using the Continuous-Capture mode will show the changes over time.

Using the Average mode will remove/reduce the effects of settling or mobile solids.

The QuadoptiX™ system measures at four points and will also reduce the effects of 

solid distribution across the sample cuvette.  

Technical Support and Warranty

For technical questions, reports of issues and any aspects related to this product, 

please contact your Palintest representative or [email protected]. Further 

contact details are found on the rear cover or on any of our websites.

When requesting technical or warranty support please quote to instrument serial 

number found on the underside of the Compact Turbimeter and in the System/

Version area of the software. The Compact Turbimeter has a two year warranty.

Calibration standards supplied has a 12 month stability. Replacement sets are 

available with part code PTC 090. Other replacement items are also available via 

your authorised representative or direct from Palintest Ltd. 

Содержание CT12

Страница 1: ...x 01428 656 808 www vanwalt com Compact Turbimeter 1 VAN WALT Monitoring your needs CONTENTS INTRODUCTION KIT CONTENTS QUADOPTIX TECHNOLOGY START UP BATTERY LIFE SYSTEM MENU READING MENU CALIBRATION M...

Страница 2: ...TENTS Compact Turbimeter Instrument CT12 Model 4 x Sample Cuvettes 4 x SDVB Calibration Standards Silicone Oil 10 ml 2 x Packs of Lint Free Cloths Sample Dilution Tube Cleaning Brush 2 x Heat Pads WP4...

Страница 3: ...ON After initialization the Mode screen will appear and display three options 1 Reading Select the mode to measure turbidity or suspended solids 2 Calibration Choose either SDVB Total Suspended Solid...

Страница 4: ...r data including the last mode used for reading Replacing the AA batteries will not lead to settings or calibration data being lost If battery power is insufficient for effective operation the Compact...

Страница 5: ...by pressing Done briefly Select OK to accept or Edit to change the entry Multiple Operator IDs can be stored and recalled from memory as required Sample ID Optionally set the Sample ID using the same...

Страница 6: ...event stray light from adversely affecting readings Microscopic scratches will refract light and can lead to higher than expected values To prevent scratches having an impact apply a thin film of sili...

Страница 7: ...ng mode select Normal and press OK Insert the sample with the orientation mark facing forward and press Read A beep will sound at the start and end of measurement Result is displayed in and stored in...

Страница 8: ...wn arrows The calibration data will be recalled automatically Insert the sample with the orientation mark facing forward and press Read A beep will occur at the start and end of each measurement cycle...

Страница 9: ...can be used 2 Factor Using data manipulated externally a slope and intercept for the straight line relationship can be entered directly Calibrating the Compact Turbimeter SDVB Select Calibration and p...

Страница 10: ...ady been created use the Up Down arrows to highlight and press OK to select If a new Sample ID is required move the highlight to a blank area and press New Create the new Sample ID using the Up Down b...

Страница 11: ...mine any offset that may have occurred due to fouling Subjecting the standards to prolonged exposure to low temperatures will reduce the NTU value The use of a WP48 Heat Pad is recommenced in these ci...

Страница 12: ...materials Readings are unstable variable Turbidity is a measurement of scattered light with the source of scattering being microscopic and larger solids present in the sample Solids can be mobile or...

Отзывы: