Using your 1200CM Series 
Microscope

Focusing adjustment is achieved by turning the coarse/fine focus controls (see 

figures 9 and 10).  The large knob is used for coarse adjustment, the smaller 

knob for fine adjustment.  The coaxial arrangement allows for easy, precise 

adjustment without stage drift.

Turning the coarse/fine focus control raises and lowers the stage vertically.  One 

complete turn of the fine focusing knob raises or lowers the stage 0.3mm; the 

smallest graduation refers to 2 microns of vertical movement.  One complete 

turn of the coarse focusing knob raises or lowers the stage 3.6mm.  To ensure 

long life, turn the focusing knobs slowly and uniformly.
The focusing tension control knob is located just inside of the right-hand 

focus  control  knob  (see  figure  9).    For  tighter  tension,  turn  the  control 

knob in a clockwise motion.  For looser tension, turn the control knob in a 

counterclockwise motion.

Vertical Focusing:  The condenser can be raised and lowered with the substage 

adjustment knob to focus the light for optimal illumination.

Aperture Adjustment:  The light path can be adjusted with the iris diaphragm 

adjustment lever located underneath the condenser.  Aperture adjustments are 

made to induce contrast into a specimen, not to adjust light intensity.
The mechanical stage X-Y controls, located underneath the right-hand side 

of  the  stage  (see  figure  9),  provide  easy  and  accurate  positioning  of  the 

sample.  One complete turn of the longitudinal (Y) control (lower half of the 

stage controls) will move the specimen 34mm left or right.  One complete turn 

of the transverse (X) control (upper half of the stage controls) will move the 

specimen 20mm front or back.
The spring-loaded slide holder can be removed for users who prefer to not 

use a mechanical stage.  Simply loosen the knurled slide holder knobs which 

lock the slide holder on the stage, and slip out the slide holder.

Variable Lighting Control/Power Switch

1

7

6

5

4

3

Figure 9

Figure 10

Focusing and Mechanical Stage Mechanisms

VanGuard 1200CM Microscopes are equipped with a rheostat controlled variable lighting control, which allows 

the user to set the lighting anywhere between off and full brightness. The variable lighting control is located on 

the left side of the microscope in front of the power switch. To increase the brightness roll the wheel towards the 

rear of the microscope.
The rocker switch located on the left hand side of the microscope (see page 6) turns the power on or off. To turn 

the power on press the rocker on the end marked “|”.  To turn the power off press the rocker on the end marked “O”.

10

All Models

All Models

Содержание 1200CM Series

Страница 1: ...Covering Models 1210CM 1211CM 1220CM 1221CM 1222CM 1223CM 1230CM 1231CM 1232CM 1233CM Operation Manual 1200CM Series...

Страница 2: ...25 N A Abbe condenser Phase contrast models come with a 1 25 N A Zernike condenser with phase annulus rings for 10X 25X 40X and 100X also has a 0 setting for brightfield work All condensers are mount...

Страница 3: ...0X Plan Phase Objective 1 ea 20X Plan Phase Objective 1 ea 40X Plan Phase Objective 1 ea 100X Plan Phase Objective 1 ea Dispersion Filter 1 ea Blue Filter 1 ea Green Filter 1 ea Model 1230CM Trinocula...

Страница 4: ...xels for 1210 and 1220 series 1200 CDPC 3 USB DIgital Camera System 3 Megapixels for 1230 series Other Accessories Part Number Description Call for Part Numbers Eyepieces 10X 16X 20X Call for Part Num...

Страница 5: ...tering Knobs Brightfield only Aperature Iris Diaphragm Adjustment Lever Fine Focus Control Substage Collector Base Stand Mechanical Stage Slide Holder Slide Holder Knobs Focusing Tension Control Knob...

Страница 6: ...Fuse Holder Parts Accessories Phase Annuli Centering Knobs Phase Contrast Condenser Iris Diaphragm Adjustment Light Bulb Centering Mechanism Substage Collector Set Screw 6 Light Bulb Access Door Lock...

Страница 7: ...eyetubes see figure 2 After removing the objectives from their storage containers individually install each one into the nosepiece by threading them clockwise into the threaded holes of the nosepiece...

Страница 8: ...er will reduce the field so that a small white hexagon is visible within a black field see Figure 6A Focusing of the hexagon is performed by turning the coarse fine focus controls This white hexagon i...

Страница 9: ...nser assembly until the two rings of light are centered upon one another see Figure 8 Once the phase rings are centered remove the phase contrast centering telescope and replace with the eyepiece The...

Страница 10: ...to induce contrast into a specimen not to adjust light intensity The mechanical stage X Y controls located underneath the right hand side of the stage see figure 9 provide easy and accurate positionin...

Страница 11: ...icroscope Interpupillary and Diopter Adjustment 1 2 3 Interpupillary adjustment the distance between eyepieces is made through a folding action The Seidentopf design allows for a folding adjustment wh...

Страница 12: ...a adapter assembly until it is firmly secured with the camera port set screw Tighten the camera port set screw Pull the prism slider completely out to divert the image to the camera port The prism sli...

Страница 13: ...Models The 100X objective which comes with this microscope must be used with immersion oil in order to maintain image quality After use the objective tip needs to be wiped clean so that no oil residu...

Страница 14: ...lose the light bulb access door and screw the knob back in to lock Return the microscope to normal operating position If necessary recenter the illuminator following the instructions on page 9 1 2 7 6...

Страница 15: ...r ordering information The same care instructions apply to all optical parts on this VanGuard Microscope including the substage collector and the condenser lenses All other parts can be cleaned with a...

Страница 16: ...can t get enough light Step Possible Cause How to Check Result Solution For More Info 1 Variable lighting control VLC is turned down too far Visually inspect VLC turned up Go to next step VLC turned...

Страница 17: ...blockage Contact dealer or VanGuard Microscopes Page 17 I can t focus on the specimen Step Possible Cause How to Check Result Solution For More Info 1 Up stop set too low Visually inspect Set too low...

Страница 18: ...25 NA 8 0mm WD 1 8mm FOV Achromatic or Plan Achromatic 20X 0 40 NA 0 50mm WD 0 9mm FOV 40X 0 65 NA 0 40mm WD 0 45mm FOV 100X oil 1 25 NA 0 25mm WD 0 18mm FOV Stage Dimensions 125mm x 135mm Stage Motio...

Страница 19: ...19 Notes...

Страница 20: ...Printed in the USA No VGMNL070319 1200CM...

Отзывы: