Responsibility of the Manufacturer
8-2
ForceTriad User’s Guide
Responsibility of the Manufacturer
Valleylab is responsible for the safety, reliability, and performance of the energy
platform only if all of the following conditions have been met:
•
Installation and setup procedures in this manual are followed.
•
Assembly, operation, readjustments, modifications, or repairs are carried out
by persons authorized by Valleylab.
•
The electrical installation of the relevant room complies with local codes and
regulatory requirements, such as IEC and BSI.
•
The equipment is used in accordance with the Valleylab instructions for use.
For warranty information, refer to
Preface
chapter in this manual.
Routine Maintenance
When should the energy platform be checked or serviced?
Valleylab recommends that the energy platform be inspected by qualified service
personnel at least twice a year. This inspection should include adjusting the
system to factory specifications.
When should the power cord be checked or replaced?
Check the power cord each time you use the energy platform or at the intervals
recommended by your institution. Replace the power cord if you find exposed
wires, cracks, frayed edges, or a damaged connector.
When should the fuses be replaced?
An internal component malfunction can damage the fuses. You may need to
replace the fuses if the system fails the self-test or if the energy platform stops
functioning, even though it is receiving power from a wall outlet. Refer to the
service manual for instructions.
Notice
Refer to the energy platform service manual for maintenance recommendations
and function and output power verification procedures.