Active Electrodes
4-12
ForceTriad User’s Guide
Accessory Port Functionality
Instruments with 8 mm pins connect directly to the accessory port on the
ForceTriad energy platform. Instruments with pin diameters less than 8 mm
require the use of an E0502-1 or E0017 Valleylab adapter. Remove the adapter
when not in use.
When a single pin electrosurgical instrument is attached to the universal
footswitching accessory port, the ForceTriad energy platform detects the
instrument and displays the Accessory Port tab on the touchscreen. The
Accessory Port tab allows the user to control the power mode and power output
level at the system interface for any footswitching instrument connected.
1.
Select the power output mode waveform by pressing the associated button.
The waveforms available in the cut mode are Pure and Blend. In the coag
mode, the waveforms available are Fulgurate or Spray.
2.
Set the power to the desired output level by pressing the green up and down
arrows. Power output is displayed in watts.
3.
Activate the cut mode by stepping on the cut or yellow pedal on the
footswitch. The cut display will illuminate yellow and a tone will sound for
the duration of the activation.
Activate the coag
mode by stepping on the coag or blue pedal on the
footswitch. The coag display will illuminate blue and a tone will sound for the
duration of the activation.