Ensure Proper Grounding
Unpacking and Installation
Force 2 Service Manual
2
-
3
Important
Ensure Proper Grounding
To ensure patient safety, the Force 2 generator must be properly grounded.
It is the user's responsibility to
ensure that the electrical installation
of the relevant room complies with
local codes and regulatory
requirements such as IEC and BSI.
The ground wire in the power cord connects to the unit chassis and
ensures that no dangerous currents flow from the cabinet in the event of
internal electrical failure. Do not use extension cords or three-prong or
two-prong electrical adapters.
Undesirable leakage currents are affected by the polarization of the
50-60 Hz input power. It is the responsibility of the user to ensure proper
polarity and grounding of the power outlets supplying power to the unit.
If required, use the equipotential ground cable supplied with the
generator. The equipotential ground connector is located on the rear panel
of the generator.
Perform a System Check
Before placing the Force 2 generator into clinical use, perform a system
check as described in Section 6,
Power Up Self-Test
.