set_acq tr_post_dur
Set post-duration samples for transient data.
Command:
set_acq tr_post_dur <samples>
<samples>
Number of samples (0 ~ 2 * ddt
µs
/ tr_decimation)
Valid state:
IDLE
set_cct interval
Set coupling check transmitter / pulser interval in seconds.
Command:
set_cct interval <interval>
<interval> = 0
Disabled
<interval> > 0
Enabled with chosen interval (0.008192 ~ 35184372088832)
<interval> = -1
Triggered with command
get_tr_snapshot
(1)
Valid state:
IDLE
(1)
set_cct interval -1
is used for time of flight measurements. When command
get_tr_snapshot
is
issued the CCT is activated and the transient recorder starts simultaneously. The time difference from start
sample to the sample of first threshold crossing is the time of flight.
set_data_log enabled
Enable/disable data logging mode. Enabling logging mode is only possible, if the data log memory is empty
(execute command
clear_data_log
before).
Command:
set_data_log enabled 0|1
Valid state:
IDLE
set_datetime
Set current date and time.
Command:
set_datetime <datetime>
<datetime>
Date and time in following format:
%Y-%m-%d %H-%M-%S.%f
,
e.g.: 2020-09-29 08:29:50.301
Valid state:
IDLE
set_filter
Set filter cutoff frequencies and order.
Command:
set_filter <hp>|none <lp>|none [<order>]