set_acq cont
Enable/disable continuous mode.
Command:
set_acq cont 0|1
Valid state:
IDLE
set_acq status_interval
Set interval of status data acquisition.
Command:
set_acq status_interval <interval>
<interval>
Interval in ms (0 ~ 2.000.000), 0: disabled
Valid state:
IDLE
set_acq tr_enabled
Enable/disable transient data acquisition.
Command:
set_acq tr_enabled 0|1
Valid state:
IDLE
set_acq tr_decimation
Set decimation factor for transient data.
Command:
set_acq tr_decimation <factor>
<factor>
Decimation factor (1 ~ 100)
1: 2 MHz
2: 1 MHz
4: 0,5 MHz
…
Valid state:
IDLE
set_acq tr_pre_trig
Set pre-trigger samples for transient data.
Command:
set_acq tr_pre_trig <samples>
<samples>
Number of samples (0 ~ 2048 / tr_decimation)
Valid state:
IDLE