6.3.13
AC Adjustments and Checks
Figure 6-2 | AC Connections
1.
Select AMPS and the 20A, 600V ranges on the 2105.
2.
Set the AC Calibrator for 600V, 10A, 100Hz, 180°, operate.
3.
Adjust RV3 for 10.00 in the current display.
Check Watts display for 6000 ±6 digits.
4.
Current source to 1A operate.
Check linearity of AMPS display ±2 digits.
2105 to the 2A range. Check for 1.000 display ±3 digits.
5.
Reduce current source to 0.1A.
Check display for .100 ±3 digits.
2105 to .2A range. Check for .1000 ±8 digits.
Increase range to 2A and input to 1 amp.
6.
Switch to VOLTS display. Check each voltage range with the corresponding inputs ±6
digits (±20 digits for the 150V display).
7.
1A, 120V inputs on the 2A, 150V ranges.
Check frequency response at 60Hz, 1000Hz and 10 kHz.
Voltage accuracy ±100 digits at 10 kHz. Current accuracy ±6 digits. Watt accuracy ±6
digits.