Chapter 3 – Specifications
Standard Current Models: Range/Resolution/Test Current/Failsafe Current
Table 1.A – Standard Current Models
Range #
0
1
2
3
4
5
6
7
8
Range
Value
2Ω
20Ω
200Ω
2kΩ
20kΩ
200kΩ
2MΩ
20MΩ
200MΩ
Resolution
100μΩ
1mΩ
10mΩ
100mΩ
1Ω
10Ω
100Ω
1kΩ
10kΩ
Test
Current
10mA
10mA
1mA
100μA
10μA
1μA
100nA
10nA
1nA
Fail-Safe
Current
[1]
16mA
16mA
1.8mA
180μA
18μA
1.8μA
180nA
18nA
1.8nA
[1] Max Value. The actual measured level is provided on the certificate of calibration.
KRC Reduced Current Models: Range/Resolution/Test Current/Failsafe Current
Table 2.B - Reduced Current Models
Range #
0
1
2
3
4
5
6
Range
Value
2Ω
20Ω
200Ω
2kΩ
20kΩ
200kΩ
2MΩ
Resolution
100μΩ
1mΩ
10mΩ
100mΩ
1Ω
10Ω
100Ω
Test
Current
5mA
5mA
500μA
50μA
5μA
500nA
50nA
Fail-Safe
Current
[1]
8mA
8mA
1.8mA
180μA
18μA
1.8μA
180nA
[1] Max Value. The actual measured level is provided on the certificate of calibration.