19
Local
COMMAND
Returns meter to local mode.
Syntax:
Local<crlf>
Returns: <crlf>
Power-on default = LOCAL mode
Notes: REMOTE mode is selected when the meter receives a valid
character (not <crlf>.
LogOn
COMMAND
Turns on continues reading mode synonymous with “Ohms?”. Note: if the
devices is already Logging, the LogOn command will return “Invalid
Command”
Syntax:
LogOn<crlf>
Returns:
Measurement Value + Multiplier every 300ms.
Example for 2Ω Range:
1.2345
Example for 20Ω Range:
12.345
Example for 200Ω Range:
123.45
Example for 2kΩ Range:
1.2345k
Example for 20kΩ Range:
12.345k
Example for 200kΩ Range:
123.45k
Example for 2MΩ Range:
1.2345M
Exam
ple for 20MΩ Range:
12.345M
Example for 200MΩ Range:
123.45M
LogOn0
COMMAND
Turns on continues reading mode synonymous with “Ohms?”. Note: if the
devices is already Logging, the LogOn0 command will return “Invalid
Command”
Syntax:
LogOn0<crlf>
Returns:
Measurement Value + Multiplier every 500ms.
Example for 2Ω Range:
1.2345
Example for 20Ω Range:
12.345
Example for 200Ω Range:
123.45
Example for 2kΩ Range:
1.2345k
Example for 20kΩ Range:
12.345k
Example for 200kΩ Range:
123.45k
Example for 2MΩ
Range:
1.2345M
Example for 20MΩ Range:
12.345M
Example for 200MΩ Range:
123.45M