background image

Section 3 | Powering the Unit 

 

 

Page | 11 

©

 2021 

– Valeport Ltd 

 

3  Powering the Unit 

3.1  Battery Warning 

 

 

Batteries are prone to gassing and in the case of Alkaline batteries 
leaking potassium hydroxide, a caustic agent that can cause 
respiratory, eye, and skin irritations. You can reduce the risks by not 
mixing battery types in the same device and by replacing all the 
batteries at the same time and prior to full discharge 

Do not mix old and new batteries or battery types. Doing so will 
reduce overall performance and may cause battery leakage or 
rupture. We recommend replacing all the batteries at the same time 

 

3.2  To Turn the TideMaster On 

The TideMaster switches on when the power button on the top right of the front panel is pressed (

quick 

press

).  The LED to the left of the panel will flash ten times in rapid succession to indicate the unit is 

initialising. 

The unit will then fall into its default sampling mode (or the last set mode).  The start of a data collection 
burst is denoted by a single flash.  The end of the burst is denoted by five flashes in quick succession.  The 
burst modes sample at fixed intervals and so it may be some time before the first burst is collected.  During 
this time the unit will be in sleep mode. 

For example: 

 

In B2 mode (6 min interval, 40 second burst) the TideMaster will collect data at: 

 

 

00/06/12/18/24/30/36/42/48/54  

 

 

minutes past the hour. 

 

 

In B4 mode (15 min interval, 40 second burst) the TideMaster will collect data at  

 

 

00/15/30/45  

 

 

minutes past the hour. 

 

3.2.1 Delayed Start 

There is no delayed start mode and the TideMaster should only be powered on when ready to collect data. 

 

3.3  To Turn the TideMaster Off 

Press and hold the Power button until the LED light comes on.  On the display version a powering off 
message will be displayed.  When the LED light goes out, the TideMaster has powered off. 

The TideMaster can only be turned off in sleep or sampling mode.   
If the TideMaster is interrupted either via PC control or via the integral display this mode will have to 
be exited before the TideMaster can be turned off 

 

 

Содержание TideMaster

Страница 1: ...t be copied reprinted or reproduced in any material form either wholly or in part and the contents of this document and any method or technique available there from must not be disclosed to any other...

Страница 2: ...evelling 9 2 3 4 Wiring 9 2 3 4 1 Junction box 9 2 4 Calibration 10 2 4 1 Pressure Sensor 10 2 4 2 VRS 20 Radar Level Sensor 10 3 Powering the Unit 11 3 1 Alkaline Battery Warning 11 3 2 To Turn the T...

Страница 3: ...or Calibration 36 8 Data Display in TideMaster Express 37 8 1 Real Time Data 37 8 1 1 File Period 37 8 1 2 Daily File Location 37 8 1 3 Backup File Location 37 8 1 4 Remote Data File Location 37 8 1 5...

Страница 4: ...g Information 51 10 1 Tide Pressure 51 10 2 Tide Radar 51 10 3 WindSONIC 52 10 4 METPAK II 52 10 5 MaxiMet 52 10 6 miniCT 53 10 7 3m Y Lead RS232 53 11 Declaration of Conformity TideMaster 54 11 1 UKC...

Страница 5: ...o Sounder system and a range of hydrographic processing software The logger is powered by 4 C type cells which together with the SD memory allows operation for over 1 year of sampling at a 15 minutes...

Страница 6: ...AL 1094978189 1 issue 1 1 2021 Valeport Ltd Page 2 1 2 Features TideMaster shown below with optional OLED control display panel Non Display model is identical other than the absence of the display and...

Страница 7: ...with 20m cable Other pressure ranges and cable lengths are available Accuracy 0 1 Full Scale Calibration Held within logging unit Dimensions 18mm diameter x 80mm VRS 20 Radar Level Sensor Type Pulsed...

Страница 8: ...nomous sampling External Power 9 28 V DC Memory 512 MB SD card memory allowing for effectively unlimited storage Sampling Raw data sampled at 8Hz and logged as average over burst 6 pre programmed burs...

Страница 9: ...e supplied screw NB This can only be done before mounting the bracket to the wall For removal of the bar it can be pushed up from the bottom of the housing until it can be reached at the top and remov...

Страница 10: ...unting requirements for the VRS 20 are a sturdy stable surface or boom that does not experience any significant vertical movement The surface does not necessarily need to be level as the VRS 20 mounti...

Страница 11: ...Section 2 Installation Page 7 2021 Valeport Ltd 2 3 1 Boom Mounting The radar is supplied with a stainless steel clamp as pictured for attaching to railings scaffold poles...

Страница 12: ...om constructed from standard scaffolding parts and attached to existing railing structure 2 3 2 Direct Mounting The clamp can be removed and the stainless steel adaptor plate fitted directly to a moun...

Страница 13: ...request The cable is glanded into the VRS 20 at the sensor end and issued as unterminated cable at the other end This is to allow easy routing of the cable through walls bulkheads etc 2 3 4 1Junction...

Страница 14: ...um for the particular site The exact calibration is affected by transducer offset water density and gravity The transducer is temperature compensated but this compensation does rely on the temperature...

Страница 15: ...ast set mode The start of a data collection burst is denoted by a single flash The end of the burst is denoted by five flashes in quick succession The burst modes sample at fixed intervals and so it m...

Страница 16: ...the control panel to reset the clock Dry the top of the housing to ensure no water enters the battery compartment during the process Undo the thumbscrews on the top of the housing They are fitted wit...

Страница 17: ...wer and prevents the unit from entering sleep mode It is recommended that unless externally powered Bluetooth connections are terminated when not required 3 5 1 Scenario A Druck Pressure Transducer Fi...

Страница 18: ...seconds per 30 minutes 182 days 3 5 4 Scenario D Druck Pressure Transducer Fitted Bluetooth On Every burst Display Period 5s Gill Windsonic Fitted Mode Pattern Battery Life B1 30 seconds per 1 minute...

Страница 19: ...splays and extract and view logged data from logging instruments Additional Features of TideMaster Express Support for up to 10 station tide gauge network Support for multiple serial inputs Cable Blue...

Страница 20: ...ected to the PC via a serial cable Before Bluetooth communications can be established it is necessary to pair the TideMaster with the PC to be communicated with A TideMaster can be paired with multipl...

Страница 21: ...pt the TideMaster and Navigate to Bluetooth menu Select pairing option Communications Bluetooth Pairing Using TideMaster Express Connect with the TideMaster via cable See Section 6 2 1 Select the Pair...

Страница 22: ...be assigned two COM port numbers the important one is the outgoing COM port as you will need to select this port when connecting to the instrument Now your TideMaster is successfully paired you will...

Страница 23: ...ns available and the inputs required Navigation through the menus is achieved using a combination of the four buttons on the front of the TideMaster Button Functions Power Button Turns unit on and off...

Страница 24: ...Factor Enter Value Site Cal Enter Low Value Enter Low Tide Value Take Pressure reading Enter High Value Enter High Tide Value Take Pressure reading Calculate Cal Output Units dBar Metres Feet Disabled...

Страница 25: ...niCT sensor 074 1 Enables miniCT sensor 6 2 2 Sampling Set Up The sampling set up of the instrument can be configured in this menu Default burst modes can be selected or a custom sampling mode can be...

Страница 26: ...p 9600 19200 38400 57600 115200 Bluetooth Setup Enable Bluetooth Disable Bluetooth Pair with Device Code 0 Telemetry Setup Set Station ID Set Delay 6 2 5 Reset option The reset option is used for a so...

Страница 27: ...th the PC Defaults to 57600 baud Radio Valeport offer a Radio Telemetry Module allowing medium range wireless communications between a PC and TideMaster See Radio Telemetry manual 0741808a for details...

Страница 28: ...data according to the sampling scenario you have set Change Baud Rate Having established communications at a particular baud rate you may wish to change this to suit your particular requirements For e...

Страница 29: ...ster has a number of predefined sampling patterns and the option for a user defined sampling pattern The TideMaster has a very accurate real time clock This can be synchronised to the PC clock at set...

Страница 30: ...ay will come on for at the end of a burst Real time output can be transmitted in a number of formats The choice of format does not affect the logged data which is always stored in the same format Tide...

Страница 31: ...t transducer position with respect to datum In order to determine this secondary calibration some on site measurements will need to be taken to establish the relationship between decibars and depth wi...

Страница 32: ...Pascal therefore 1 dBar 104 Pascal and Density Acceleration due to Gravity 1 is the Gain Factor 7 1 2 1 Gain Factor Fresh Water Using a standard density of 994 kg m3 pure water at 15 C and a gravity f...

Страница 33: ...set should be entered as a positive value To check the operation of the unit with these user input values select cal check to view data from the transducer in both pressure dBar and Depth m ft formats...

Страница 34: ...t is possible to mimic a tidal cycle by raising the transducer by a known amount This is described below as the moving transducer method Site Calibrations can be carried out using either of two method...

Страница 35: ...thod Install and fix the transducer in the desired position and connect it to the TideMaster Run the TideMaster Express program on the PC Connect the TideMaster to the PC having previously selected th...

Страница 36: ...tion option Select Site Calibration and required output units Low Calibration At the lowest expected tide Enter the Low Tide Height either read from a calibrated Tide Board or staffing gauge relative...

Страница 37: ...either order and repeated if necessary It is possible to disconnect from the TideMaster and return later to carry out the second part of the calibration Once you are happy with the collected pressure...

Страница 38: ...relative to a known datum Press the Get Reading button TideMaster will sample pressure readings for 30s and record the high height pressure relationship Move the transducer up by a known distance i e...

Страница 39: ...is factory calibrated and the calibration is held within the radar sensor This calibration is independent of environmental conditions To transform this range to a tide height the offset between the R...

Страница 40: ...ollow either the moving or fixed transducer method entering Low High values using the appropriate menu option when prompted When you are happy with the values entered use the Calculate Calibration opt...

Страница 41: ...ter Express will monitor its last used serial port and log any data received on that port This data will be logged and backed up The File menu provides access to configuration of file periods and loca...

Страница 42: ...her analysis or manipulation A confirmation window will appear to show that the process is complete 8 3 Data Display TideMaster Express Simple A numerical display large characters of the latest readin...

Страница 43: ...Display will immediately indicate the last received reading for that parameter 8 3 2 Scroll Display This display type gives a sequential list of data points from all available parameters When opened...

Страница 44: ...be quickly shown of any logged data file without adding it the Sensor List The Y axis of the graph will auto scale for each parameter to cover the range of data currently being displayed The resolutio...

Страница 45: ...ing a numerical indication of the magnitude and bearing This display will appear when Wind Rose Display is selected from the Display Type menu and will be updated as new data is received The display m...

Страница 46: ...ch will show the data for a full calendar day from midnight to midnight or 24 Hour Band which shows a moving graph of data for a full 24 hour period centred on the current time i e 12 hours before and...

Страница 47: ...sure 10s interval Radar B1 Burst sampling pattern of 30 seconds 1 minute B2 Burst sampling pattern of 40 seconds 6 minutes B3 Burst sampling pattern of 40 seconds 10 minutes B4 Burst sampling pattern...

Страница 48: ...t Firmware version 0741705A5 File Creation Date 23 01 2013 14 27 23 Battery Level 5 7 TideMaster S N 3 Station ID 01 Site info Valeport test site Calibrated 18 07 2012 Mode B1 Pressure units m output...

Страница 49: ...Level 5 7 TideMaster S N 3 Station ID 01 Site info Valeport test site Calibrated 18 07 2012 Mode B1 Pressure units m output format TIDEMASTER User Pressure cal Gain 1 234000 Offset 1 000000 Vale Pres...

Страница 50: ...units m output format TIDEMASTER Radar Min limit 0 500 Max Limit 2 000 Datum 2 000 Pre amble 4 Wind speed units Knots Air Pressure units mBar Air Temperature units C Time stamp TAB Depth TAB Depth SD...

Страница 51: ...2013 14 33 36 Battery Level 5 7 TideMaster S N 3 Station ID 01 Site info Valeport test site Calibrated 18 07 2012 Mode B1 Pressure units m output format TIDEMASTER Radar Min limit 0 500 Max Limit 20...

Страница 52: ...chars ID Station ID Date ddmmyyyy Time hhmmss Tide Height tt ttt Tide height stdev h hhh Tide height units D 0 1 2 dBar Metres Feet Battery voltage vv v Status hh checksum of the above CL cr lf end of...

Страница 53: ...20 11 2009 09 30 00 P dBar 3 472 D m 3 136 CR LF Where ID 01 Station ID 1 10 TIME 20 11 2009 09 28 00 Time Date DD MM YYY hh mm ss P dBar 3 487 Tide Height in Decibars D m 3 152 Tide Height in Metres...

Страница 54: ...e ID 01 TIME 20 11 2009 09 36 00 TIDE HEIGHT 3 073 ID 01 TIME 20 11 2009 09 37 00 TIDE HEIGHT 3 064 ID 01 TIME 20 11 2009 09 38 00 TIDE HEIGHT 3 052 ID 01 TIME 20 11 2009 09 39 00 TIDE HEIGHT 3 042 ID...

Страница 55: ...dified UTS UTS6JC10E6P A V B GND C SIG D SIG E N C F Screen 10 2 Tide Radar END 1 0745518 PCB Into Radar Junction Box END 2 0745518 PCB Out of Radar Junction Box END 3 UTS6JC10E7S TideMaster END 4 D T...

Страница 56: ...ONNECTOR PIN CONNECTOR PIN J5 8 way Push fit Block Pin 1 is nearest edge of board adjacent to earth post 8 6 WAY UTS UTS6JC10E6P A V Supply to Gill 7 B GND Supply to Gill 4 C RS232 Out to Gill 3 D RS2...

Страница 57: ...pply to MiniCT 3 C RS232 Out to MiniCT 2 D RS232 In from MiniCT 1 E MiniCT 232 Gnd F Screen 10 7 3m Y Lead RS232 END 1 7 Way UTS Connector red ring END 2 9 WAY D TYPE END 3 BANANA PLUGS FUNCTION CONNE...

Страница 58: ...horised by Valeport Ltd could void the CE compliance of the product and negate your authority to operate it This product has demonstrated CE compliance under conditions that include the use of shielde...

Страница 59: ...Section 11 Declaration of Conformity TideMaster Page 55 2021 Valeport Ltd 11 2 EU CE Certificate...