Valeport miniSERIES Series Скачать руководство пользователя страница 18

© 2019 Valeport Ltd

0660880d -  miniCTD, miniSVP and miniTIDE

Page 18

6.  Battery Life

6.1.  miniSVP & miniCTD

The units draw power continuously whilst running.  Battery life is calculated as follows:

Alkaline cells have a nominal capacity of 6000mAh.  
We assume 77% efficiency, giving a usable life of 4620mAh.  
With a 1.5v alkaline C cell, the current drain is 120mA.
4620mAh / 120mA = 38½  hours

Lithium cells have a nominal capacity of 7700mAh.
We assume 66% efficiency, giving a usable life of 5082mAh.
With a 3.6v lithium C cell, the current drain is 50mA.
5082mAh / 50mA = 101  hours (approximately)

6.1.1.  Operation with Bluetooth

Using a standard C-cell battery, the MiniSVP/CTD fitted with a Bluetooth logger should have an endurance of
approximately 12 hours continuous operation.  Once enabled (mag key removed) the Bluetooth unit is in
continuous transmit mode, even underwater.

6.2.  miniTIDE

In continuous output mode, the instrument draws power continuously whilst running:

Alkaline cells have a nominal capacity of 6000mAh.
We assume 77% efficiency, giving a usable life of 4620mAh.
With a 1.5v alkaline C cell, the current drain is 62mA.
4620mAh / 62mA = 7 4½  hours

Lithium cells have a nominal capacity of 7700mAh.
We assume 66% efficiency, giving a usable life of 5082mAh.
With a 3.6v alkaline C cell, the current drain is 24mA.
5082mAh / 24mA = 212  hours (approximately)

In Burst mode, the unit cycles between ‘Run’ and ‘Sleep’ modes, making calculations more complex.  The table
summarises the battery life in each of the preset patterns:

Mode

Pattern

Alkaline

Lithium

B1

30 seconds per 1 minute

6 days

17 days

B2

40 seconds per 5 minutes

21 days

46 days

B3

40 seconds per 10 minutes

38 days

67 days

B4

40 seconds per 15 minutes

52 days

80 days

B5

60 seconds per 30 minutes

64 days

87 days

Note that all battery calculation figures are estimated, and may vary according to deployment temperature
and the inherent battery variability.  Valeport accepts no liability for the failure of a battery to last for the
expected lifetime.

Содержание miniSERIES Series

Страница 1: ...roduced in any material form either wholly or in part and the contents of this document and any method or technique available therefrom must not be disclosed to any other person whatsoever without the...

Страница 2: ...ry 10 3 6 Output 10 3 7 Operating Modes 11 3 8 Performance 12 4 Data Requests and Output Formats 12 4 1 Stop Command 12 4 2 Baud Rates 12 4 3 Run Commands miniSVP miniCTD only 13 4 4 Run Commands mini...

Страница 3: ...er 20 7 1 Bluetooth miniLogger Pack 21 7 2 USB Bluetooth Receiver 22 8 Care and Maintenance 22 8 1 SubConn Care 22 8 2 Greasing and mating above water dry mate 23 9 O Rings 24 10 Wiring Information 24...

Страница 4: ...uthorised by Valeport Ltd could void the CE compliance of the product and negate your authority to operate it This product has demonstrated CE compliance under conditions that include the use of shiel...

Страница 5: ...2019 Valeport Ltd EU Declaration of Conformity CE Marking Page 5 1 1 CE Marking Standard Equipment...

Страница 6: ...2019 Valeport Ltd 0660880d miniCTD miniSVP and miniTIDE Page 6 1 2 CE Marking Bluetooth Equipment...

Страница 7: ...2019 Valeport Ltd EU Declaration of Conformity CE Marking Page 7...

Страница 8: ...g modes to suit profiling and moored applications Each product in the range is available in either 500m rated Acetal or 6000m rated titanium housing This manual highlights where this may result in spe...

Страница 9: ...housings Wiring Information is in Section 6 Materials Part mini SVP mini CTD mini TIDE Material Main housing Titanium 6000m or Acetal 500m Sensor bulkhead Titanium 6000m or Acetal 500m Space Rods Carb...

Страница 10: ...strument is fitted with SD memory cards This provides more than sufficient capacity for the operational life of the battery pack 3 6 Output Units are fitted with both RS232 and half duplex RS485 commu...

Страница 11: ...me taken for a pulse of sound to travel a known distance The material selected does not measurably expand over the operating temperatures of the instrument ensuring the highest possible accuracy at al...

Страница 12: ...ception of the Stop command All commands are echoed back by the instrument as they are typed 4 1 Stop Command The instrument can be stopped at any time by sending the character The instrument returns...

Страница 13: ...implified UNESCO formula 018 2 Pressure is output in feet calculated using simplified UNESCO formula 4 6 Pressure Tare 009 Takes a single reading of the pressure sensor and uses the measured value as...

Страница 14: ...e g RN File1 MyData FREE Outputs amount of free memory VOLT Reads internal and external voltage levels 4 10 Data Formats Each time the instrument is turned on or a new file started in Burst mode on th...

Страница 15: ...PPPP P e g 0123 4 dBar PPP PP e g 012 34 dBar PP PPP e g 12 345 dBar For the miniTIDE pressure data is a fixed length string 3 decimal places including leading zeroes The temperature data miniSVP mini...

Страница 16: ...affected To compensate for this an adjustment will have to be made to the calibration parameters Contact Valeport for further details 5 1 Cage The deployment cage is a rugged deployment method providi...

Страница 17: ...deployment wire or rope and also for a small weight to be added below the instrument to alleviate streaming Whichever deployment system is used it is important to note that if the data is being monit...

Страница 18: ...e the instrument draws power continuously whilst running Alkaline cells have a nominal capacity of 6000mAh We assume 77 efficiency giving a usable life of 4620mAh With a 1 5v alkaline C cell the curre...

Страница 19: ...ry fitted then please use DataLog x2 software or the codes defined in Section 3 9 to reset the clock Disconnect any external connections including the switch plug from the instrument s SubConn connect...

Страница 20: ...the standard logger pack On first installation it will require 2 minutes initialise during this period the magnetic switch key should be removed After this period of initialisation the switch key can...

Страница 21: ...and should install automatically on most windows PC s If the receiver will not automatically install drivers can be downloaded from http www ftdichip com Drivers VCP htm Once installed the device will...

Страница 22: ...ommended lubricant is Loctite 8021 in a spray can or Molykote 44 Medium but use very sparingly Any accumulation of sand or mud in the female contact should be removed with fresh water Failure to do so...

Страница 23: ...pear to be damaged in any way or have become hard due to age Lightly greased not twisted and completely clean before reassembly Ordering Information The main O ring fitted to the Logger Module all ins...

Страница 24: ...2 Rx From PC 9 RS232 Ground 10 Internal Battery Enable Link to RS232 Ground Mode Pins Pins 5 and 6 on the output connector are designated as Mode Pins By grounding a combination of these pins the inst...

Страница 25: ...7 NC 8 NC 9 NC 10 Link to Pin 1 10 3 3m Y Lead RS232 3m Cable 10 Way SubConn 1m Power Cable 4mm Banana Plugs 1m Data Cable 9 Way D Type Pin Pin Function 1 Power Ground BLACK plug 2 Power V RED plug 3...

Отзывы: