Section 6 | Frequently Asked Questions
©
2022
–
Valeport Ltd
Page | 20
How fast can I profile?
The profiler runs at 8Hz, and in free fall will achieve a drop rate of perhaps 2m/s. Falling at this
rate, it will cover 0.25m between readings, so provided the profile is set to take a reading at a
greater increment than this, it will function correctly.
The SV Sensor Outputs Zero Sometimes
–
Why Is That?
The sensor outputs zero when it doesn’t record the returning sound pulse within the expected time
frame (a time frame that equates to 1400
–
1600m/s in terms of sound velocity). The most
common occurrence of a zero value is when the sensor is in air, but it can also happen if the probe
has been dropped into a soft bed and is covered in mud or sediment. This will normally wash off
during the up-cast. It can also happen if the sensor has been deployed for some time without
cleaning, and there is significant growth on the sensor.