4. Maintenance
4.1 Periodic maintenance
Every 1-2 years, make sure that the printed circuit board is not corroded.
4.1.1 Testing proper operation
For easy testing and adjustment, reference signals are provided for the 2 output channels. For
instance, by inserting a jumper plug, an output can be forced to the full scale or zero level. A
readout device can be easily adjusted to comply with the transmitter output. Inside the
transmitter box, there is a label showing all the useful jumper combinations for output scale,
measurement range, and testing.
Table 6 Jumper selectable reference sources for testing
Variable
Reference signal alternatives
Direction
000°, 008°, 357°
Speed
0.0 m/s, 0.8 m/s, 51.2 m/s
1. To remove WAT12 cover, remove the 4 screws holding the cover.
2. To enable the test stage, remove the jumper
J5
from the jumper block X4.
3. Select one of the direction and speed alternatives with jumpers
J1
,
J2
,
J3
, and
J4
.
Table 7 Output values for jumpers in test mode
Jumper
Output value
J1 ON
51.2 m/s
J2 ON
0.8 m/s
J3 ON
357°
J4 ON
008°
All open
0 m/s and 000°
For the location of the jumper block X4, see
4. Verify the readings with the data collecting system at the receiving end.
5. To end the test stage, reattach the jumper
J5
to the jumper block X4.
6. After a successful test, carefully reattach the enclosure cover with the 4 screws.
Chapter 4 – Maintenance
27
Содержание WAT12
Страница 1: ...M210309EN B User Guide Analog Wind Transmitter WAT12...
Страница 6: ...WAT12 User Guide M210309EN B 4...
Страница 23: ...Figure 15 Wiring with WHP151 mains power supply Chapter 3 Installation 21...
Страница 24: ...Figure 16 Wiring with WHP25 mains power supply and WA25 sensors WAT12 User Guide M210309EN B 22...
Страница 34: ......
Страница 36: ......
Страница 37: ......
Страница 38: ...www vaisala com...