User's Guide _______________________________________________________________________
62____________________________________________________________________ M210543EN-B
Automatic Message Sending
In the automatic (CLOSEd) mode PWD22 sends the predefined
message at selected intervals. Automatic message is selected with the
AMES command.
AMES Message_number Message_interval
where
Message_number =
Valid range 0 ... 4 and 7. Selects the
corresponding message. Any negative
message number is converted to 0. If only the
message number is given, the previous
message interval setting is used.
Message_interval =
Given in multiples of 15 s (= measuring
interval). Therefore, intervals 15, 30, 45... are
valid. Other intervals are converted to
multiples of 15 s. Maximum sending interval
is 255 s (4 min 15 s).
See the example below:
AMES 0 60
Selects message number 0 to be sent once in a minute.
Messages can also be displayed in the command mode with the MES
command, described in section MES on page 66 .
Message Polling
In the polled (CLOSEd) mode, PWD22 sends a data message when
the host computer transmits a polling command. The message polling
mode is selected with the following command:
AMES Message_number 0
where
Message_number =
Valid range 0 ... 4 and 7. Selects the
corresponding message as the default polled
message. Any negative message number is
converted to 0.