PRODUCT DESCRIPTION

The PTB220TS and a PTB220AxB2Ax barometer (order specified) are
assembled together at factory to become a Transfer/travelling standard.
It is also possible to assemble the customer's existing PTB220
barometer and separately ordered wooden case (PTB220CASE); see the
instructions on page 5. Please note, that there are certain requirements
for the PTB220 barometer to be compatible with the case. The
barometer should have the following features:

 

1, 2 or 3 pressure transducers

 

class A accuracy

 

local display

 

RS232/485 serial interface

This guide gives the basic operating instructions of the PTB220TS. It
does not, however, replace the operating manual of the PTB220 series
digital barometers. For detailed operation instructions and descriptions
refer to the User's Guide of the PTB220.

The PTB220TS includes the following parts

 

a wooden (oak) carrying case

 

wide and comfortable carrying handle

 

two key locks

 

a space reserved for small parts and documents in the case lid

 

tilting frame where the barometer is mounted for table-top use

 

incorporated sealed lead acid battery (YUASA NP0.8-12)

 

power cord for recharging from car cigarette lighter socket

 

DC plug for a local mains adaptor cord

 

pressure connectors: for 1/8" tubing and a 5 mm connector

 

9-pin sub D-connector (male)

 

6 pin 5 mm screw-terminal bus (female)

Содержание PTB220TS / CASE

Страница 1: ...PTB220TS CASE USER S GUIDE U370EN 1 1 JUNE 1999...

Страница 2: ...may be reproduced in any form or by any means electronic or mechanical including photocopying nor may its contents be communicated to a third party without prior written permission of the copyright ho...

Страница 3: ...1 OPERATION 2 Operating voltage 2 Tilting frame 3 Battery 3 RS232 3 RS485 4 LED Indicators 4 ASSEMBLING THE CASE AND AN EXISTING BAROMETER 5 TECHNICAL DATA 7 Electronics 7 Electromagnetic compatibilit...

Страница 4: ...essure transducers class A accuracy local display RS232 485 serial interface This guide gives the basic operating instructions of the PTB220TS It does not however replace the operating manual of the P...

Страница 5: ...closed if the pressure connector is detached For measurement the connector MUST be on place Operating voltage DC voltage of 10 25 VDC is required to operate the barometer A power cord with a suitable...

Страница 6: ...removed from the case by lifting upwards NOTE The pressure connector has to be removed before closing the wooden carrying case Battery The battery gives a twelve hour service period The period is even...

Страница 7: ...r of the PTB220TS CASE is as following PIN SIGNAL 1 not used 2 RS485 HI 3 not used 4 not used 5 RS485 LO 6 not used LED Indicators There are four LED indicators in the back plate of the tilting frame...

Страница 8: ...asily be detached after the barometer is installed into the case 1 Remove the wooden tilting frame from the case 2 Unscrew the base plate 3 Attach the PTB220 onto the base with four M6x16 screws 4 Thr...

Страница 9: ...ering Place the conducting wire end to left hand side 9 Attach the spacers 2 pcs M3x12 Fix the support bar of the pressure port to the spacers 10 Attach the M3x18 spacers 4 pcs 11 Connect the flat cab...

Страница 10: ...f duplex RS485 half duplex Connectors 9 pin female sub D connector 6 pin 5 mm screw terminal bus Electromagnetic compatibility PTB220TS CASE with the PTB220 barometer is fully compatible according to...

Отзывы: