background image

ASSEMBL

ING THE CASE 

AND AN 

EXISTING BAROMETER

_________________________________________

V

AISALA

______________________________________________________________________________5

ASSEMBLING THE CASE AND AN
EXISTING BAROMETER

To assemble the existing PTB220 digital barometer and the wooden
case follow the instructions below.

NOTE! It is recommended to remove the cover hinge of the barometer
by loosening its screws. Then the cover can easily be detached after the
barometer is installed into the case.

1. Remove the wooden tilting frame from

the case.

2. Unscrew the base plate.
3. Attach the PTB220 onto the base with

four M6x16 screws.

4. Thread the flat cable between the

battery covering and the base plate.
Connect the cable to D-terminal.

5. Remove the original 1/8" pressure

fitting of the barometer. Then, attach

the pressure tubing connector to the

PTB220. It is recommended to use
Loctite 572 joint paste for sealing.

6. Place the assembly into the tilting

frame. Check that the flat cable and
the pressure tubing come out of the

back opening.

7.

 

Attach the base with six screws

.

All manuals and user guides at all-guides.com

Содержание PTB220Case

Страница 1: ...PTB220TS CASE USER S GUIDE U370EN 1 1 JUNE 1999 All manuals and user guides at all guides com a l l g u i d e s c o m...

Страница 2: ...y form or by any means electronic or mechanical including photocopying nor may its contents be communicated to a third party without prior written permission of the copyright holder The contents of in...

Страница 3: ...voltage 2 Tilting frame 3 Battery 3 RS232 3 RS485 4 LED Indicators 4 ASSEMBLING THE CASE AND AN EXISTING BAROMETER 5 TECHNICAL DATA 7 Electronics 7 Electromagnetic compatibility 7 Battery 7 Housing 7...

Страница 4: ...s A accuracy local display RS232 485 serial interface This guide gives the basic operating instructions of the PTB220TS It does not however replace the operating manual of the PTB220 series digital ba...

Страница 5: ...connector is detached For measurement the connector MUST be on place Operating voltage DC voltage of 10 25 VDC is required to operate the barometer A power cord with a suitable plug in connector is s...

Страница 6: ...wards NOTE The pressure connector has to be removed before closing the wooden carrying case Battery The battery gives a twelve hour service period The period is even longer if the back light is not us...

Страница 7: ...is as following PIN SIGNAL 1 not used 2 RS485 HI 3 not used 4 not used 5 RS485 LO 6 not used LED Indicators There are four LED indicators in the back plate of the tilting frame The LED indicators info...

Страница 8: ...the barometer is installed into the case 1 Remove the wooden tilting frame from the case 2 Unscrew the base plate 3 Attach the PTB220 onto the base with four M6x16 screws 4 Thread the flat cable betw...

Страница 9: ...ing wire end to left hand side 9 Attach the spacers 2 pcs M3x12 Fix the support bar of the pressure port to the spacers 10 Attach the M3x18 spacers 4 pcs 11 Connect the flat cable and the battery cond...

Страница 10: ...lex Connectors 9 pin female sub D connector 6 pin 5 mm screw terminal bus Electromagnetic compatibility PTB220TS CASE with the PTB220 barometer is fully compatible according to standards EN50081 1 EN...

Отзывы: