200
PR-23 instruction manual
If you are using Windows (or Mac OS X 10.3 or newer or any recent Linux distrib-
ution) and the computer has the default network settings, change the
DTR IP ad-
dress
to 169.254.x.y, where x=1–254 and y=1–254, for example 169.254.100.100 or
169.254.123.1. This way the DTR address will be suitably paired with the address
your computer automatically generates for itself.
If in doubt, you can get your Windows computer’s network settings by opening the
command window (command prompt) and by typing the command
ipconfig
at the
command prompt (press Enter to give the command), see figure 12.7 (in Mac OS X and
Linux the same command is called
ifconfig
). The result will give you your computer’s
IP address, so you can change the DTR to match; the connection should always work
if you match the first three groups of numbers and just change the last number.
Figure 12.7
Typical IP configuration for a stand-alone laptop
when connected to a DTR; laptop wireless (WLAN) is turned off
Note:
You may have to connect the cross-over cable and power on the DTR before
your computer generates an IP address for the ethernet connection (computer reboot
may also be required). Note also that the connection will not work if the computer
and the DTR have exactly the same IP address.
Note:
Please make sure that your WLAN (Wireless network connection) is not active
when you connect to the DTR. If the WLAN is active, the computer’s Ethernet connec-
tion may not function as expected.
When you have set the DTR (and/or the computer) according to instructions above,
you can proceed to test the connection as instructed below in Section 12.3.
12.3 Testing the Ethernet connection
On the Ethernet connector inside the DTR there are two
diagnostic LEDs
. The
green
LED indicates that the physical connection is working
, i.e. that both ends of the
Ethernet cable are plugged in, the device in each end is powered and the cable is of
correct type. The
orange LED indicates traffic in the cable
, i.e. that DTR receives
data.
The IP address can be tested with a
ping
command after the physical Ethernet connec-
tion has been set up and the DTR is powered. In Windows systems
ping
is available by
using the Command Prompt (usually found in the Accessories). The usage of
ping
is
Содержание K-PATENTS PR-23 Series
Страница 1: ...IM EN PR23 v 2 00 Instruction Manual Vaisala K PATENTS Process Refractometer PR 23...
Страница 2: ......
Страница 4: ......
Страница 17: ...2 Inline refractometer sensor 7 2 2 2 PR 23 mounting guide...
Страница 26: ...16 PR 23 instruction manual...
Страница 29: ...4 Prism wash systems 19 Figure 4 1 A prism wash system for steam non sanitary...
Страница 30: ...20 PR 23 instruction manual Figure 4 2 A sanitary prism wash system for steam...
Страница 32: ...22 PR 23 instruction manual Figure 4 5 Wiring for a prism wash system for steam...
Страница 33: ...4 Prism wash systems 23 Figure 4 6 A prism wash system for high pressure water non sanitary...
Страница 34: ...24 PR 23 instruction manual Figure 4 7 A sanitary prism wash system for high pressure water...
Страница 35: ...4 Prism wash systems 25 Figure 4 8 Wiring for a prism wash system for high pressure water...
Страница 39: ...4 Prism wash systems 29 Figure 4 12 Mounting of the wash nozzle for Process refractometer PR 23 GP...
Страница 40: ...30 PR 23 instruction manual...
Страница 50: ...40 PR 23 instruction manual...
Страница 72: ...62 PR 23 instruction manual...
Страница 74: ...64 PR 23 instruction manual Figure 8 2 Motherboard PR 10600 and H1 interface card PR 10701 in detail...
Страница 113: ...9 Sensor specifications 103 WZ WZ WZ WZ WZ Figure 9 15 Mounting the sensor with a wfc flow cell...
Страница 151: ...9 Sensor specifications 141 Figure 9 38 Intrinsically safe wiring PR 23 IA according to WRG 362...
Страница 152: ...142 PR 23 instruction manual Figure 9 39 Intrinsically safe wiring PR 23 IF with STR according to WRG 478...
Страница 174: ...164 PR 23 instruction manual...
Страница 179: ...11 Safe Drive 169 11 3 3 Safe Drive steam wash system parts...
Страница 186: ...176 PR 23 instruction manual 11 4 2 Wiring Figure 11 9 PR 23 SD system wiring...
Страница 204: ...194 PR 23 instruction manual Figure 11 19 Markings on the retractor handle...
Страница 206: ...196 PR 23 instruction manual...
Страница 218: ...208 PR 23 instruction manual...
Страница 226: ...216 PR 23 instruction manual...
Страница 230: ...220 PR 23 instruction manual...
Страница 232: ...222...
Страница 238: ...228...
Страница 240: ...230...
Страница 242: ...232...
Страница 244: ...234...
Страница 246: ...www vaisala com...