9 Sensor specifications
101
9.5.4 PR-23-GC mounting specifics
The Compact refractometer is mounted either in a pipe elbow by a Sandvik coupling
connection or in a straight pipe via a Wafer flow cell or a Pipe flow cell. Both flow
cell mounting designs create optimum flow velocity on the measurement surface pro-
viding a good self-cleaning effect. The Wafer flow cell has also an optional automatic
wash system capability.
In a pipe at least 2.5" in diameter the sensor is mounted in a pipe elbow with a Sand-
vik coupling. In a 2" pipe the sensor is mounted in a pipe elbow via reducing ferrule
PR-9283. In 0.5, 1 and 1.5 inch pipes a flow cell is installed in a straight pipe. The
flangeless Wafer flow cell is a compact alternative to traditional flow cells. The Wafer
refers to a flow cell body that is installed between DIN, ANSI or JIS piping flanges with
bolts and nuts. The Wafer flow cell is a one-piece body construction with no welds. A
Pipe flow cell is also available for 0.5 and 1 inch pipes.
$
$
6(&7,21$$
Figure 9.12
Mounting the sensor in a pipe 2 1/2" or larger
Содержание K-PATENTS PR-23 Series
Страница 1: ...IM EN PR23 v 2 00 Instruction Manual Vaisala K PATENTS Process Refractometer PR 23...
Страница 2: ......
Страница 4: ......
Страница 17: ...2 Inline refractometer sensor 7 2 2 2 PR 23 mounting guide...
Страница 26: ...16 PR 23 instruction manual...
Страница 29: ...4 Prism wash systems 19 Figure 4 1 A prism wash system for steam non sanitary...
Страница 30: ...20 PR 23 instruction manual Figure 4 2 A sanitary prism wash system for steam...
Страница 32: ...22 PR 23 instruction manual Figure 4 5 Wiring for a prism wash system for steam...
Страница 33: ...4 Prism wash systems 23 Figure 4 6 A prism wash system for high pressure water non sanitary...
Страница 34: ...24 PR 23 instruction manual Figure 4 7 A sanitary prism wash system for high pressure water...
Страница 35: ...4 Prism wash systems 25 Figure 4 8 Wiring for a prism wash system for high pressure water...
Страница 39: ...4 Prism wash systems 29 Figure 4 12 Mounting of the wash nozzle for Process refractometer PR 23 GP...
Страница 40: ...30 PR 23 instruction manual...
Страница 50: ...40 PR 23 instruction manual...
Страница 72: ...62 PR 23 instruction manual...
Страница 74: ...64 PR 23 instruction manual Figure 8 2 Motherboard PR 10600 and H1 interface card PR 10701 in detail...
Страница 113: ...9 Sensor specifications 103 WZ WZ WZ WZ WZ Figure 9 15 Mounting the sensor with a wfc flow cell...
Страница 151: ...9 Sensor specifications 141 Figure 9 38 Intrinsically safe wiring PR 23 IA according to WRG 362...
Страница 152: ...142 PR 23 instruction manual Figure 9 39 Intrinsically safe wiring PR 23 IF with STR according to WRG 478...
Страница 174: ...164 PR 23 instruction manual...
Страница 179: ...11 Safe Drive 169 11 3 3 Safe Drive steam wash system parts...
Страница 186: ...176 PR 23 instruction manual 11 4 2 Wiring Figure 11 9 PR 23 SD system wiring...
Страница 204: ...194 PR 23 instruction manual Figure 11 19 Markings on the retractor handle...
Страница 206: ...196 PR 23 instruction manual...
Страница 218: ...208 PR 23 instruction manual...
Страница 226: ...216 PR 23 instruction manual...
Страница 230: ...220 PR 23 instruction manual...
Страница 232: ...222...
Страница 238: ...228...
Страница 240: ...230...
Страница 242: ...232...
Страница 244: ...234...
Страница 246: ...www vaisala com...