Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 37
High end adjustement
6.
After having made the low end adjustment, insert the probe head
into a measurement hole of the high end reference chamber (e.g.
NaCl: 75 % RH chamber in the humidity calibrator HMK15, please
use the adapter fitting (13.5 mm)). Please, note that the difference
between the two humidity references must be at least 50% RH.
7.
Let the probe stabilize at least 30 minutes. You can follow the
stabilization by pressing any key, typing C and pressing
ENTER
.
8.
When stabilized, type the high end reference value after a question
mark and press
ENTER
.
>crh
RH : 11.25 Ref1 ? c
RH : 11.24 Ref1 ? c
RH : 11.24 Ref1 ? 11.3
Press any key when ready ...
RH : 75.45 Ref2 ? c
RH : 75.57 Ref2 ? c
RH : 75.55 Ref2 ? c
RH : 75.59 Ref2 ? 75.5
OK
>
OK
indicates that the calibration has succeeded
9.
Take the probe out of the reference conditions and replace the
filter. Take care to tighten the filter properly, recommended force
130 Ncm.
10.
If needed, give the calibration information (date and text) to the
transmitter's memory, see the serial commands on page 29.
11.
Reset the transmitter by giving a command
RESET
. Transmitter