Chapter 3 _________________________________________________________________Mounting
VAISALA _______________________________________________________________________ 17
NOTE
It is recommended that the supply is switched off, before connecting the
cable plug to the transmitter during installation or service/calibration.
Mounting the MMT317
The MMT317 has a small pressure-tight probe and it is ideal for tight
spaces with threaded connection. The small probe is installed using
the threaded fitting bodies, see below.
NOTE
Do not unsolder and then again resolder the sensor head cable from
and to the printed board during installation; this procedure may alter
the humidity calibration of the transmitter.
MMT317 with Swagelok Connector
for Tight-place Installations
Swagelok installation kit for the MMT317 includes Swagelok
connector with ISO3/8" thread ( Vaisala order code: SWG12ISO38)
or NPT1/2" thread (Vaisala order code: SWG12NPT12). See figure
below for MMT317 with Swagelok Installation Kit.