User's Guide ______________________________________________________________________
170 _________________________________________________________________ M210566EN-J
Inputs and Outputs
Operating voltage
10 ... 35 VDC, 24 VAC
with optional power supply module
100 ... 240 VAC, 50/60 Hz
Start-up time after power-up
3 s
Power consumption at 20 °C (Uin 24VDC)
RS-232
max 25 mA
Uout 2 × 0 ... 1V / 0 ... 5V / 0 ... 10V
max 25 mA
Iout 2 × 0 ... 20 mA
max 60 mA
display and backlight
+ 20 mA
during sensor purge
+ 110 mA max
Analog outputs (2 standard, 3rd optional)
current output
0 ... 20 mA, 4 ... 20 mA
voltage output
0 ... 1 V, 0 ... 5 V, 0 ... 10 V
Accuracy of analog outputs at 20 °C
± 0.05 % full scale
Temperature dependence of the analog outputs
± 0.005 %/°C full scale
External loads
current outputs
RL < 500 ohm
0 ... 1V output
RL > 2 kohm
0 ... 5V and 0 ... 10V outputs
RL > 10 kohm
Max wire size
0.5 mm2 (AWG 20)
stranded wires recommended
Digital outputs
RS-232
RS-422/485
(optional)
LAN
(optional)
WLAN
(optional)
Protocols
ASCII commands
MODBUS
RTU
MODBUS
TCP
Relay outputs (optional)
0.5 A, 250 VAC, SPDT
Display (optional)
LCD with backlight, graphic
trend display
Menu languages
Chinese, English, Finnish,
French, German, Japanese,
Russian, Spanish, Swedish
Mechanics
Cable bushing
M20x1.5 for cable diameter
8 ... 11mm/0.31..0.43"
Conduit fitting
1/2"NPT
User cable connector (optional)
M12 series 8- pin (male)
option 1
with plug (female) with 5 m /
16.4 ft black cable
option 2
with plug (female) with screw
terminals
Probe cable diameter
HMT333
80ºC
6.0
mm
Other
probes
5.5
mm
Probe cable lengths
2 m, 5 m or 10 m
Probe tube material
HMT331
Chromed ABS plastic
Other probes
AISI 316L
Housing material
G-AlSi 10 Mg (DIN 1725)
Housing classification
IP 65 (NEMA 4)