User's Guide ______________________________________________________________________
160 _________________________________________________________________ M210566EN-J
RH : 75.45 Ref2 ?
c
RH : 75.57 Ref2 ?
c
RH : 75.55 Ref2 ?
c
RH : 75.59 Ref2 ?
75.5
OK
>
12.
OK
indicates that the adjustment has succeeded and the new
calibration coefficients are calculated and stored. Enter the
adjustment information (date and text) to the memory of the
transmitter; see the commands
CTEXT
and
CDATE
.
13. Press
the
ADJ
button on the motherboard to close the adjustment
mode.
14. Take the probe out of the reference conditions and replace the
filter.
Relative Humidity Adjustment after Sensor
Change
Using Display/Keypad
When using the optional display/keypad, follow the instructions in
section Using Display/Keypad on page 158 but select
Adj. for new RH
sensor (
instead of
1-point
/
2-point adjustment)
.
Using Serial Line
After sensor change, carry out the procedure as described in previous
sections. Just replace the
CRH
command with the
FCRH
command.
FCRH
Example:
>
fcrh
RH :
1.82
1.
ref
?
0
Press any key when ready...
RH
:
74.22
2. ref
?
75
OK
>
The
OK
indicates that the calibration has succeeded.