CHAPTER 8
TECHNICAL DATA
This chapter provides the technical data of the product.
Specifications
Table 11
Relative Humidity Specifications
Property
Description / Value
Measuring range
0 ... 100 %RH
Accuracy (including nonlinearity,
hysteresis and repeatability)
with
Vaisala HUMICAP®180
for typical applications
Vaisala HUMICAP®180R
for typical applications
Vaisala HUMICAP®180C
for applications with chemical purge
and/or warmed probe
Vaisala HUMICAP®180RC
for applications with chemical purge
and/or warmed probe
at +15 ... 25 °C
± 1 % RH (0 ... 90 %RH)
± 1.7 % RH (90 ... 100 %RH)
at -20 ... +40 °C
± (1.0 + 0.008 × reading) %RH
at -40 ... +180 °C
± (1.5 + 0.015 × reading) %RH
with
Vaisala HUMICAP®180L2
for applications with demanding
chemical environment
at -10 ... +40 °C
± (1.0 + 0.01 × reading) %RH
at -40 ... +180 °C
± (1.5 + 0.02 × reading) %RH
Factory calibration uncertainty
(+20 °C)
±0.6 % RH (0 ... 40 %RH)
±1.0 % RH (40 ... 97 %RH)
(Defined as ± 2 standard deviation
limits. Small variations possible, see
also calibration certificate.)
Response time (90 %) for
HUMICAP®180 and
HUMICAP®180C at 20 °C in still air
8 s with grid filter
20 s with grid + steel netting filter
40 s with sintered filter