HMI41 AND PROBES
M210316EN-A
Operating Manual
47
Accuracy of wet bulb temperature (°F)
RH/%
10
20
30
40
50
60
70
80
90
100
T/
°
F
-40
0.36
0.36
0.36
0.36
0.36
0.36
0.36
0.37
-
-
-4
0.37
0.38
0.38
0.38
0.39
0.39
0.40
0.40
-
-
32
0.49
0.50
0.51
0.51
0.52
0.53
0.54
0.55
0.55
0.56
68
0.82
0.81
0.81
0.80
0.79
0.78
0.78
0.77
0.76
0.76
104
1.51
1.39
1.29
1.21
1.15
1.09
1.05
1.00
0.97
0.94
140
2.62
2.16
1.86
1.64
1.49
1.37
1.28
1.20
1.14
1.09
176
4.03
2.96
2.38
2.03
1.79
1.61
1.48
1.38
1.29
1.22
212
5.52
3.68
2.84
2.36
2.05
1.83
1.66
1.54
1.44
1.36
248
6.94
4.33
3.26
2.67
2.30
2.04
1.85
1.70
1.59
1.49
284
8.23
4.92
3.65
2.97
2.54
2.25
2.03
1.87
1.74
1.63
320
9.41
5.48
4.03
3.26
2.78
2.45
2.21
2.03
1.89
1.77
11.5
Electromagnetic compatibility
Complies with the following standard:
EN 61326-1:1997 +Am 1:1998, Electrical equipment for measurement,
control and laboratory use - EMC requirements: Light environment.