User’s Guide ________________________________________________ HM40
34
Calibrating the HM40 Using the
HMK15 Humidity Calibrator
The HMK15 Humidity calibrator allows you to produce
known humidity environments using saturated salt
solutions.
Performing a good calibration takes some time
and preparation. Read the HMK15 User’s Guide
before performing your first calibration with the
HMK15.
1.
Press the Menu button and
select the Calibration submenu.
2.
Select the quantity to be
calibrated at menu item [1]
Quantity. You can calibrate
Temperature (T) or Relative
Humidity (RH) measurement.
All other humidity quantities are
calculated from RH and T, so
they will also be adjusted.