Serial interface _____________________________________________________________________
Vaisala _________________________________________________________________________ 41
The transmitter asks and measures relative humidity and calculates the
calibration coefficients, for example:
>CRH
↵
RH
:
1.82 1. ref
?
0
↵
Press any key when ready...
↵
RH
:
74.22
2. ref
? 75
↵
OK
>
The
OK
indicates that the calibration has succeeded.
Temperature adjustment
CT
↵
The transmitter asks and measures temperature readings and calculates
the calibration coefficients, for example:
>CT
↵
T
:
0.81
1. ref
?
0.5
↵
Press any key when ready...
↵
T
: 99.12
2.
ref
?
99.5
↵
OK
>
The
OK
indicates that the calibration has succeeded. When performing
one-point calibration press only
↵
for 2. reference.
Analog output calibration
Connect the HMT360 to a multimeter. Give the following command.
ACAL
↵
1.
Disconnect the serial cable from the transmitter while reading the
multimeter value for the CH1 (I1).
2.
Reconnect the serial cable. Type the multimeter reading and press
↵
.
3.
Type the higher current multimeter reading and press
↵
.
>ACAL
↵
Ch1 I1
(mA) ?
4.846
↵
Ch1 I2
(mA) ?
19.987
↵
>