Integer Values (x100, Metric)
261
0104
hex
Absolute humidity
16-bit signed
integer
g/m
3
262
0105
hex
Mixing ratio
16-bit signed
integer
g/kg
263
0106
hex
Wet-bulb temperature
16-bit signed
integer
°C
264
0107
hex
Enthalpy
16-bit signed
integer
kJ/kg
Integer Values (x100, Non-metric)
385
0180
hex
Relative humidity
16-bit signed
integer
%RH
386
0181
hex
Temperature
16-bit signed
integer
°F
387
0182
hex
Dew point temperature
16-bit signed
integer
°F
388
0183
hex
Dew/frost point
temperature
16-bit signed
integer
°F
389
0184
hex
Absolute humidity
16-bit signed
integer
gr/ft
3
390
0185
hex
Mixing ratio
16-bit signed
integer
gr/lb
391
0186
hex
Wet-bulb temperature
16-bit signed
integer
°F
392
0187
hex
Enthalpy
16-bit signed
integer
Btu/lb
1)
NOTE: 16-bit integers have a maximum value of +32767. Certain measurement parameters
(for example, mixing ratio and enthalpy) can exceed this value in x100 scaling. In such cases,
the value of the parameter is cut off at +32767, and measurements above this value are not
reported. Verify that the measurement ranges in your application are suitable for the 16-bit
integer format with x100 scaling; whenever possible, it is recommended to use 32-bit float
values.
20
M212264EN-B
Содержание HMD60 Series
Страница 4: ...4 M212264EN B ...
Страница 22: ...22 M212264EN B ...
Страница 94: ...94 M212264EN B ...
Страница 97: ...Запрещается прикасаться к чувствительному элементу ОСТОРОЖНО 97 РУССКИЙ ...
Страница 116: ...116 M212264EN B ...
Страница 134: ...134 M212264EN B ...
Страница 152: ...152 M212264EN B ...
Страница 153: ......
Страница 154: ...www vaisala com ...