HMI41 Calibration Option

19165ZZ-U220en-1.1

Operating Manual

1996-08-29

3

2.

CALIBRATION

For a successful calibration, it is essential that the probe of the HMI41 and
that of the transmitter are at the same temperature, and that the reference
probe has been previously calibrated. Always allow enough time for the
readings to stabilize. Note that the stabilization time depends on the ambient
conditions and may vary from 10 minutes to a couple of hours.

2.1

Getting started

After having selected the desired function (Chapter 1.2), turn the HMI41 off
and connect the calibration cable to the EXT connector at the bottom of the
HMI41 (Figure 2.1). Connect the other end of the cable to the appropriate
connector in the transmitter (Figures 2.2 - 2.4).

                               

EN TER

MO DE

HOLD

ON/OFF

EXT

C o n n e c t o r   f o r   t h e

c a l i b r a t i o n   c a b l e

Figure 2.1

Location of the HMI41 calibration connector

one point
RH calibration

test connector

Figure 2.2

Calibration connector and potentiometer in                    

the HMP130 series transmitters

Содержание HMD20

Страница 1: ...USER S GUIDE Calibration of Series HMD W20 30 and HMP130 Transmitters with Vaisala HUMICAP Indicator HMI41 U220EN11...

Страница 2: ...echanical including photocopying nor may its contents be communicated to a third party without prior written permission of the copyright holder The contents are subject to change without prior notice...

Страница 3: ...20en 1 1 Operating Manual 1996 08 29 i Contents 1 INTRODUCTION 1 1 1 HMI41 with calibration option 1 1 2 Selecting the HMI41 calibrator function 1 2 CALIBRATION 3 2 1 Getting started 3 2 2 Calibration...

Страница 4: ...HMI41 Calibration Option Operating Manual 19165ZZ U220en 1 1 ii 1996 08 29 This page intentionally left blank...

Страница 5: ...libration options please refer to corresponding manuals Calibration is quick and easy to perform In calibration the HMI41 measures relative humidity with the reference probe and receives the RH inform...

Страница 6: ...librator for HMD W 60 70 and HMP140 series analogue transmitters 3 HMI41 as a calibrator for HMP230 series digital transmitters 4 HMI41 as a calibrator for the HMD W20 30 and HMP130 series analogue tr...

Страница 7: ...may vary from 10 minutes to a couple of hours 2 1 Getting started After having selected the desired function Chapter 1 2 turn the HMI41 off and connect the calibration cable to the EXT connector at t...

Страница 8: ...h the ON OFF button and wait until a text similar to the following appears on the display RH Numbers on the first line indicate the RH measured by the transmitter and numbers on the second line indica...

Страница 9: ...ference probe If you wish you can now calibrate the transmitter by pressing MODE again the measured values appear on the display Wait until the readings have stabilized and adjust the calibration pote...

Страница 10: ...unction press ON OFF until you can see some text on the display Then release the ON OFF button and press within 1 2 seconds both ENTER and MODE buttons until the text setup appears on the display Then...

Страница 11: ...ge RH Let the readings stabilize and then adjust the transmitter potentiometers until the readings are the same offset 65 RH gain 65 RH Alternatively you can adjust the transmitter when the display sh...

Страница 12: ......

Страница 13: ......

Страница 14: ...www vaisala com U220EN...

Отзывы: