1.3.2 Configuring Relays with GMP250 Software Version 1.2.9
Use the
rsel
command (not the
asel
command) to set the analog output channel to
current (mA).
1. Connect GMP251/GMP252 to a computer with USB service cable 242659 and open a
terminal connection (for example, PuTTy).
2. Enable advanced configuration commands with the
pass 1300
command.
3. Set the serial mode of the analog current output channel (2) to relay control with the
following command:
smode relay2
4. Configure the current output channel's relay settings with the
rsel
command:
rsel [channel parameter release_ppm set_ppm set_value error_value]
•
channel
= analog output channel: set to
2
(current)
•
parameter
= the parameter that controls the relay: set to
co2
•
release_ppm
= CO
2
ppm level at which the relay is released
•
set_ppm
= CO
2
ppm level at which the relay is activated
•
set_value
= the current output (mA) when the relay is activated: set to
12
•
error_value
= the current output when the probe is in error state
The relay mode of GMP251 and GMP252 probes has been designed to use 12 mA as
the relay control current. Use only 12 mA for the
set_value
configuration when
using the relay box with GMP251 and GMP252.
Example:
Set the analog output channel 2 (current) to control relay operation with the following
settings:
• The relay is activated when the measurement is over 10100 ppm CO
2
and released
when under 9900 ppm CO
2
. Setting these relay activation limits means using a
1 % CO
2
(10 000 ppm CO
2
) relay setpoint with a hysteresis of ±1 % (100 ppm CO
2
).
• The output current is 12 mA when the relay is activated.
• The output current is 0 mA when the probe is in error state.
Enter the following commands:
pass 1300
smode relay2
rsel 2 co2 9900 10100 12 0
Chapter 1 – Relay Box Quick Guide
7