1.3.1 Configuring Relays with GMP250 Software Version 1.3.1 or
Later
Use the
rsel
command (not the
asel
command) to set the analog output channel to
current (mA).
1. Connect GMP251/GMP252 to a computer with USB service cable 242659 and open a
terminal connection (for example, PuTTy).
2. Enable advanced configuration commands with the
pass 1300
command.
3. Set the serial mode of the analog current output channel (2) to relay control with the
following command:
smode relay2
Chapter 1 – Relay Box Quick Guide
5