Chapter 4 _________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 115
The precipitation measurement can be calibrated by comparing the
PWD32 rain amount to measurements made with a suitable reference
rain gauge.
This comparison should be made after a few rain events with 5 mm or
more of total accumulated rain. A new scaling factor can be calculated
using the following formula:
NEWSCALE = OLDSCALE * (REF_AMOUNT / PWD_AMOUNT)
where
OLDSCALE
= The old value of rain intensity scale.
REF_AMOUNT = The amount measured with the reference rain
gauge.
PWD_AMOUNT = The corresponding amount measured by
PWD32.
CLRS
The
CLRS
command resets the precipitation accumulations to 0.00 and
000.
This can be done by the host with the following command:
;
PW
id
C
C
R
where
;
= ESC
(
ASCII character 1B hex)
C
R
= ASCII character (13 hex)
System Configuration Commands
The following command can be used to display system parameters and to
edit the current system configuration:
-
PAR
, Parameter message
-
CONF
password
, Updates configuration
-
BAUD
rate comm. type
, Sets baud rate and type (Rate 300,
1200, 2400, 4800, 9600), (
Communication type
E (7E1) or N
(8N1)