Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 71
Command
Description
CHEC
Displays average signal
CAL
Calibrator
_
frequency
Calibration
CLRS
Clears precipitation sums
ACAL
Analog output calibration
TIME
hh mm ss
Sets/displays system time
DATE
yyyy mm dd
Sets/displays system date
BAUD
rate
par
Baud rate setting
(
Rate
300, 1200, 4800, 9600)
(
Par
E(7E1) or N(8N1)
AN
channel
Analog channel
(0,1,3,8 ... 15 or ANALOG ID)
DAC
data
(Without DATA = SWEEP)
RESET
Hardware reset by watchdog
WHIS
Present weather history
WSET
PRW reference values
DRY ON
Sets DRD dry offset
WET ON
Sets DRD wet scale
BLSC
Sets/Displays background luminance scale
MES Command
After opening the line for operator commands (see section
Entering/Exiting the Command Mode on page 61), a data message can
be displayed using the MES command. There are eight messages
available for different uses and they numbered from 0 to 7. Refer to
section Message Types on page 63 for message type descriptions.
The command format is the following:
MES
Message_number
↵
↵↵
↵
with a valid range from 0 to 7. For example, when choosing the data
message number 2, type
MES 2
↵
AMES Command
The AMES command defines the message, which the FD12P
transmits as the automatic message or as the default polled message.
Messages can also be displayed by the MES
command, described in