User's Guide _______________________________________________________________________
58
______________________________________________________ M210296en-A
Local practice may require changes, especially in the precipitation
intensity limits and the haze threshold value. For details, see the
description of the WSET command in section WSET Command on
page 73.
When the precipitation intensity and amount measurement is not
factory calibrated, higher accuracy can be achieved by adjusting a
scaling factor with the WSET command. The new scaling factor can
be calculated by comparing the FD12P against a reference rain gauge.
For details, see the description of the WSET command in section
WSET Command on page 73.