Vaetrix ETG Series User Manual
•
Admin – Adjust Calibration
The unit will prompt you to apply the high reference
pressure. We recommend that you are within 10% of the
high pressure reference setpoint. Enter in the known value
by using the ARROW keys to increase or decrease digits and
confirm with the SELECT key.
Follow the same steps for the low pressure reference
setpoint. Once again, we recommend you are with 10% of
the setpoint.
•
Admin – Adjust Calibration
The unit will respond by with “Storing Calibration” and
return to Measure Mode.
Certain pressure ranges may have more than two pressure
reference setpoints. Follow the same procedure of entering
in the known value and confirming with SELECT key. Once
the calibration is complete verify the gauge is within
specification by the running the up and down scale
throughout range. Reference the original Certificate of
Calibration for cardinal points.