OPERATION

The eGauge is initiated by pushing the ON/OFF key. 

When the button is pushed there will be a “Snap” 

or tactile response, no further force is necessary to 

stimulate the gauge to respond. When the gauge is 

turned on the screen shown to the right will appear 

on the eGauge. The initialize screen will show the 

current date and time and current firmware revision. 

The second screen will display 

Full Scale Pressure

eGauge Serial Number

Last Calibration Date

and 

Current Update Rate

.

When “MENUS” is displayed, two menus are displayed 

as the screen below shows.

The two menu’s available are 

Pressure Units 

Enabled

 and the 

Gauge Update Rate

. To select 

a menu, use the “NEXT” and “PREV” buttons to 

navigate to the desired menu. Once the menu has 

been selected, press “SEL” to display it. The screen 

to the right shows the Pressure Units which are 

enabled. To navigate to a unit, use the “NEXT” and 

“PREV” buttons. When a unit is selected it will flash. 

The “SEL” button will toggle the value either “Y” or 

“N”. When all of the desired units have been enabled 

or disabled, press the “EXIT” button to return to the 

top menu. The default from the factory is all units 

are enabled.

4

Содержание e Gauge 1

Страница 1: ...SINCE1958 OPERATION MANUAL for Vaetrix PLOT Application eGauge 1 eGauge 3...

Страница 2: ...ric Pressure 10 Status Indicators 10 Low Battery Indicator 10 Error Codes 10 Communication 11 Changing the Batteries 11 Recalibration 11 Applications 13 Product Specifications 15 Troubleshooting 17 Er...

Страница 3: ...housed in aluminum bodies and will be hereafter referred to as eGauge A MIN MAX function captures the extremes of varying pressure signals The data logging function allows the capability to pause chan...

Страница 4: ...pdate Rate When MENUS is displayed two menus are displayed as the screen below shows The two menu s available are Pressure Units Enabled and the Gauge Update Rate To select a menu use the NEXT and PRE...

Страница 5: ...the process the gauge is The display will briefly indicate the full scale pressure of installed transducer serial number date of last calibration current date and time and the barometric pressure rea...

Страница 6: ...ions are defined depending on the operating mode of the eGauge and are shown on the display above each key ON OFF Button has one function and purpose to turn the eGauge on or off When the eGauge is tu...

Страница 7: ...t temperature are monitored in real time the display is updated four times per second 2 Fluid temperature is acquired through the RTD every 30 seconds and the display is updated at the time of acquisi...

Страница 8: ...be changed before logging starts 1 Txxx SEC is the time interval between samples The xxx is the number of seconds between samples 2 START button will start the logging process 3 GAUGE button will sto...

Страница 9: ...ensure enough storage space is available when gauge is logging Static pressure measurement is taken from the transducer mounted in the eGauge This very special transducer is the only repairable press...

Страница 10: ...ric transmitter If the port is obstructed DO NOT attempt to clear with high pressure air Blowing into the barometric vent will cause damage to the transmitter When cleaning the eGauge take special car...

Страница 11: ...se to open the battery compartment The cover is hinged at the bottom and should swing open freely Remove the batteries from the holders eGauge1 On replacing the batteries in the eGauge1 use a Phillips...

Страница 12: ...ed to be adjusted to the dead weight test units 4 Verify the system has NO LEAKS 5 Effect of system head is corrected by theAUTO ZERO function of the eGauge The evaluation performed will confirm calib...

Страница 13: ...ssure gauges transmitters or static pressure on chart recording systems The system will accept the use of hydraulic fluid distilled water or air dependent on type of stable pressure source used Althou...

Страница 14: ...ystem for the calibration of differential transmitters and gauges The system will accept the use of hydraulic fluid distilled water or air dependent on type of stable pressure source used Note The low...

Страница 15: ...onlinearity hysteresis repeatability and temperature effects on both span and zero drift for pressures greater than 10 of Full Scale For pressures 1 to 10 of Full Scale the accuracy is 02 of Full Scal...

Страница 16: ...inless Steel Optional For use with fluid compatible with Inconel X750 Stainless Steel CONNECTION Pressure 1 4 MNPT up to 10 000 PSI 1 4 High pressure autoclave up to 50 000 PSI Temperature 1 2 Thermo...

Страница 17: ...iminish as the gas dissolves into the fluid t Solution Evacuate all tubing with a vacuum pump before introducing fluid into the system Non repeatability of Pressure Measurements When checking the gaug...

Страница 18: ...ult factory Readings are not correct Reason Batteries are bad t Solution Replace all 4 AA Batteries After Replacing Batteries readings are still not correct t Solution Check gauge against standard tha...

Страница 19: ...ients have been corrupted in any way Err 05 is displayed t Solution Non Field repairable consult factory Err LB Displayed Reason Low Battery t Solution Need to replace Batteries Err AD Displayed Reaso...

Страница 20: ...adjust settings in the eGauge1 or eGauge3 Blue tooth SETUP The user will need to pair the eGauge Blue tooth with the PC Blue tooth This can be done through a Blue tooth wizard The following steps desc...

Страница 21: ...aph Customer Data This item is enabled when a log file has been opened It will allow the user to incorporate User s data into the graph s cover page The user also has the capability to define 3 fields...

Страница 22: ...s feature is used to maximize battery life This function is automatically disabled in the Log Mode The DAMPING function can be enabled or disabled If there is check mark beside Damping then it is enab...

Страница 23: ...23 NOTES...

Страница 24: ...For technical support concerning the operation of the Vaetrix PLOT contact our Vaetrix Brand Department at Email sales vaetrix com CONTACT FOR TECHNICAL SUPPORT AND SERVICE...

Отзывы: