Troubleshooting
20901118_EN-US_PC301x VARIO select Serie_V1.3_100720
47
Error
Cause
Remedy
Person-
nel
If necessary, use a
different, external
venting valve.
Venting valve does
not operate
Venting valve in
sensor defective.
Replace defective
components.
Specialist
Vacuum pump does
not start
Pumping unit
switched off.
Power plug not cor-
rectly plugged in or
pulled out.
VACUU·BUS plug-
in connection or ca-
bles defective or
not connected.
Overpressure in
outlet line.
Switch on Pumping
unit.
Check power con-
nection and cable.
Check
VACUU·BUS plug-
in connection and
cables to the con-
troller.
Open outlet line.
Ensure clear pas-
sage.
Operator
Vacuum pump
stopped
Vacuum pump does
not start
Motor overloaded.
Motor overheated.
Thermal protection
triggered.
Check coolant con-
nection.
Ensure supply of
coolant.
Allow the motor to
cool down.
Clear error manu-
ally:
-> Disconnect
pumping unit from
power supply
-> Eliminate cause
of error
-> Restart pumping
unit
Specialist
No or very little suc-
tion power
Leak in the suction
line or apparatus.
Condensate collec-
tion flask not
mounted correctly.
Condensate inside
the vacuum pump.
Gas ballast open
Gas ballast cap
porous or no longer
present.
Check suction line
and apparatus for
leaks.
Check condensate
collection flask and
mount correctly.
Check equipment
for leaks.
Allow vacuum
pump to run for a
few minutes with
the suction nozzle
open.
Close the gas bal-
last
Operator