background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. Manual-no.: 999021 / 07/05/2008

page 6 of 22

Ensure that in case of failure the pump and the vacuum system always will turn into a
safe status.

In case of diaphragm cracks or leaks in the manifold pumped substances might be
released into the environment or into the pump housing. To reduce the risk of leaks,
ask for a diaphragm pump with additional safety diaphragm.

Comply especially with  notes on operation and use and maintenance.

Use only 

genuine spare parts and accessories.

Otherwise safety and performance of the equipment as well as the electromagnetic
compatibility of the equipment might be reduced. Possibly the CE mark or the C/US
conformity becomes void if not using genuine spare parts.

Ensure that maintenance is done only by suitably trained and supervised technicians.
Ensure that the maintenance technician is familiar with the safety procedures which
relate to the product processed by the vacuum system and that the equipment, if
necessary, is appropriately decontaminated before starting maintenance.
Comply with local and national safety regulations.

Wear parts have to be replaced regularly. 

In case of normal wear the lifetime of the

diaphragms and valves is > 10000 operating hours. Bearings have a typical durability
of 40000 h. Motor capacitors have a typical durability in the range of 10000 to 40000 h
depending strongly on the operation conditions like ambient temperature, humidity or load.

Check every capacitor regularly by measuring its capacity and estimating its opera-
tion time. Exchange old capacitors early enough to prevent a failure. If an overaged
motor capacitor fails it might get hot and even melt and may cause a flame to form
which could be 

dangerous for persons and equipment in the vicinity

. The ca-

pacitors have to be replaced by an electrician.

Before starting 

maintenance

 vent the system, isolate the pump and other compo-

nents from the vacuum system and the electrical supply, drain condensate and allow
sufficient cooling of the pump.
Before starting maintenance, wait 

two minutes

 after isolating the equipment from

mains to allow the capacitors to discharge.

In order to comply with law (occupational, health and safety regulations, safety at work
law and regulations for environmental protection) vacuum pumps, components and
measuring instruments returned to the manufacturer can be repaired only when certain
procedures (see section 

”Notes on return to the factory”

) are followed.

The motor is shut down by a 

thermal cutout

 in the winding.

Attention: Manual reset is necessary. Switch off the pump or isolate the equipment
from mains. Wait approx. five minutes before restarting the pump.

Attention

: In case of 

supply voltage below 100V

, the lock of the cutout might be

restricted and the pump might restart on its own after sufficient cooling down. Take
suitable precautions, if an automatic restart of the pump may lead to a critical
dangerous situation.

Содержание ME 8S

Страница 1: ...the users responsibility to check carefully the validity of this document with respect to his product Manual no 696261 999021 08 05 2008 Technology for Vacuum Systems Instructions for use Customized...

Страница 2: ...re only to be used and distributed completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 696261 999021...

Страница 3: ...e customer s requirements It is the customers responsibility to check the suitability of the equipment for the in tended application prior to use The instructions for use of the standard product are e...

Страница 4: ...re only to be used and distributed completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 696261 999021...

Страница 5: ...nd unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999021 07 05 2008 page 1 of 22 ME 8 MZ 4 MD 4 MV 2 Diaphragm...

Страница 6: ...pump VACUUBRAND diaphragm pumps are the result of many years of experience in construction and practical operation of these pumps combined with the latest developments in material and manufacturing t...

Страница 7: ...duct Manual no 999021 07 05 2008 page 3 of 22 Attention Important notes Not permitted Misuse may cause damage Caution Hot surface Isolate equipment from mains Note Contents Safety information 4 Techni...

Страница 8: ...l shock The supply cable may be fitted with a moulded European IEC plug or a plug suitable for your local electrical supply If the plug has been removed or has to be removed the cable will contain wir...

Страница 9: ...ping substances which may form deposits inside the pump If there is a danger of the formation of deposits in the pump chamber check inlet and outlet of the pump inspect the pump chamber regularly and...

Страница 10: ...or capacitors have a typical durability in the range of 10000 to 40000 h depending strongly on the operation conditions like ambient temperature humidity or load Check every capacitor regularly by mea...

Страница 11: ...z H 0 6 0 5 d e e p s d a o l o N n i m 1 0 0 8 1 0 0 5 1 0 0 8 1 0 0 5 1 0 0 8 1 0 0 5 1 0 0 8 1 0 0 5 1 t a t n e r r u c d e t a R z H 0 6 0 5 V 0 0 1 z H 0 6 V 0 2 1 z H 0 6 0 5 V 8 0 2 0 0 2 z H...

Страница 12: ...a m g a r h p a i d M P F e v l a v M P F 2 V M 4 D M g n i r O M P F e l z z o n e s o h l e e t s s s e l n i a t s e g n a l f l l a m s l e e t s s s e l n i a t s r e c n e l i s E P A P e s o h...

Страница 13: ...s strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999021 07 05 2008 page 9 of 22 ME 8 MZ 4 inlet outlet with silencer pump rati...

Страница 14: ...anged It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999021 07 05 2008 page 10 of 22 MD 4 MV 2 inlet pump rating plate o...

Страница 15: ...y Switch off the pump or isolate the equipment from mains Wait approx five minutes before restarting the pump Attention In case of supply voltage below 100V the lock of the cutout might be restricted...

Страница 16: ...o systems at different pressure levels stainless steel FFKM leak rate 10 3 mbar x l s for pres sure differences 500 mbar Hose caoutchouc 10 mm ID 686002 Digital vacuum gauge DVR 2 682902 ME 8 Small fl...

Страница 17: ...ment with respect to his product Manual no 999021 07 05 2008 page 13 of 22 Assemble control valve first then align and fasten gauge ME 8 Vacuum adjustment device inlet 696840 An additional part is nec...

Страница 18: ...o c t o n g n i r g n i r e t n e C e h t n i k a e l r o d e n o i t i s o p m u u c a v r o e n i l e p i p m e t s y s a h t i w p m u p k c e h C t a y l t c e r i d e g u a g m u u c a v k c e h...

Страница 19: ...eposits mainte nance should be carried out frequently Regular maintenance will improve the lifetime of the pump and also protect both man and environment Before starting maintenance vent the system is...

Страница 20: ...using a spiky or sharp edged tool e g screw driver we recommend to use a rubber mallet or compressed air to be blown carefully into port Remove head cover carefully from housing cover Note po sition o...

Страница 21: ...pport disc Check for washers Do not mix the washers from the differ ent heads Make sure that the original number is reassem bled at the individual pump head Smaller number of washers The pump will not...

Страница 22: ...re the pres sure at the inlet port A vacuum of less than 110 mbar should be indicated If the reading is higher recheck the pump chamber and make sure that the valves and the diaphragms are correctly s...

Страница 23: ...y product returned for repair We submit quotations only on request and always at the customer s expense If an order is given the costs incurred are offset from the costs for repair or from the pur cha...

Страница 24: ...for the returned product that neither toxic corrosive biologically active explosive radio active nor contamination dangerous in any way has occurred the product is free of dangerous substances the oi...

Страница 25: ...e nous d clarons que le dispositif d sign ci dessus est conforme aux prescriptions de base des directives EU applicables et indiqu s en ci que concerne conception dessin et mod le vendu par nous m mes...

Страница 26: ...s obtained under test conditions for specific applications depend on a number of factors beyond our control It is therefore strictly the users responsibility to very carefully check the validity of ap...

Отзывы: