page 32 of 51
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999073 / 15/12/2009
Setting the
lower pressure value
(condition for automatic shut
down):
➨
Press p
▲
or p
▼
.
☞
”Set p” and the lower pressure value are displayed for approx.
1s.
☞
With a second tip within one second or continuous press-
ing:
➨
Set the lower pressure value by pressing key p
▲
or p
▼
.
☞
The lower pressure value is stored as the new set point
when the key is released.
Note:
It is recommended to select a pressure value which is
approx. 10 mbar higher than the ultimate pressure value achiev-
able in the system.
Setting the process parameters
Setting the
time for automatic shut down
:
➨
Press key
▲
or
▼
.
☞
”Set”, the clock symbol and the time for automatic shut down
are displayed for approx. 1 s.
☞
With a second tip within one second or continuous press-
ing:
➨
Press key
▲
or
▼
to set the value for the time for shut down
(1 - 200 min).
☞
The value for the shut down time is stored as the new set
point when the key is released.
☞
If the actual pressure falls below the lower set point, the
time for automatic shut down starts to run. The time for au-
tomatic shut down is reset if the actual pressure rises higher
than the lower set point (with a hysteresis of 3 mbar).
The process control stops when the time for shut down has
elapsed. The vacuum pump is switched off and the coolant valve
is closed.
☞
Pump starts again in case of great pressure increase.
☞
If the actual pressure is rising above the upper set point, the
pump starts again (corresponds to starting by pressing key
START/STOP).
☞
The process control can be stopped manually at any time
by pressing key START/STOP.
Setting the
upper pressure value
(condition for restart):
➨
Keep key MODE pressed and press key
▲
additionally.
☞
”Set p”, ”arrow up” and the upper pressure value are dis-
played for approx. one second.
247
mbar
67
mbar
MODE
67
mbar
p
015
☞
With a second tip within one second or continuous press-
ing:
➨
Set the upper pressure value by pressing key
▲
or
▼
.
☞
The upper pressure value is stored as the new set point
when the key is released.