page 26 of 51
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999073 / 15/12/2009
Automatic mode
Switching on the automatic mode
➨
Press key MODE.
☞
”Auto” is displayed.
☞
After starting process control, the pressure is adapted to
the process automatically: The controller determines the
boiling point and adapts if the boiling pressure changes.
☞
If automatic mode is switched on while process control is
active, the process pressure is adapted.
Setting the maximum speed:
Set the value for the speed by using key
▲
or
▼
(
effective
only if the process pressure is adapted
):
☞
To activate set mode: Press key shortly. The current speed
is displayed for one second.
☞
With a second tip within one second or continuous press-
ing:
☞
Set speed. The new set point is stored when the key is re-
leased.
☞
The reduced maximum speed restricts the evaporation rate
in case of low boiling solvents and a carry over of sub-
stances.
☞
The utilisation of a condenser or a cold trap at the inlet can
be optimised.
☞
☞
☞
☞
☞
If selecting ”HI” (recommended) for motor speed, the
maximum speed for automatic adaptation of the motor
speed is restricted to a value corresponding to the sys-
tem size.
10.0
MODE
MODE
MODE
If
automatic switching off
was preselected and the controller
has determined that evaporation is completely finished:
☞
”Auto” is flashing.
☞
The pump is switched off. If the controller measures a pres-
sure > 1060 mbar, the pressure is reduced to 1060 mbar to
avoid overpressure.
☞
The controller is still active, the control must be finished by
pressing key START/STOP or by venting (the coolant valve
will not be closed until then and until the after-running time
has elapsed).
To use the advantages of the automatic mode optimally, switch
on ”Auto” before starting the process.
☞
Switching on automatic mode: Press key MODE.
☞
The controller determines the operating pressure.
☞
When activating the automatic mode if process control is
active (after having pressed key START/STOP) the actual
pressure is stored as set point and adjusted automatically.
Switching off the automatic mode
➨
Press key MODE.
☞
Switching off the automatic mode is possible at any time.
☞
”Auto” disappears.
☞
While vacuum control is active the actual pressure is stored
as new set point.