Changing the selection at the voltage se-
lection switch:
•
Change the selection at the voltage se-
lection switch only, if the pump is un-
plugged from the power source.
1.
Disconnect the electrical power cord.
2. Use a screw driver to adjust the voltage se-
lection switch to the supply voltage:
”115/120” corresponds to 100-120 V and
”230/240” corresponds to 200-230 V.
MD 1 VARIO-SP (DC motor)
• Check the power source and the pump’s
rating plate to be sure that the power source
and the equipment match in voltage, phase,
and frequency.
Connect the supply voltage and the control
signal to the pump’s control line (see „Control-
ling the MD 1 VARIO-SP“, pg. 38).
Switch the supply voltage and, if required, the
control signal on.
Содержание MD 1
Страница 1: ...page 1 of 72 Diaphragm pumps Instructions for use Technology for Vacuum Systems MD 1 VARIO SP MD 1 ...
Страница 27: ...page 27 of 72 MD 1 MD 1 C US pump with dual voltage motor 4 9 8 7 5 6 3 2 1 2 1 2 1 ...
Страница 28: ...page 28 of 72 MD 1 VARIO SP Ensure sufficient venting of the pump 9 5 6 7 10 ...
Страница 53: ...page 53 of 72 1 2 1 2 3 15 mm 4 6x size 2 2x 1 ...
Страница 54: ...page 54 of 72 4 5 6 20696264 Clean ...
Страница 55: ...page 55 of 72 Replacing the diaphragm 7 8 9 1 2 ...
Страница 56: ...page 56 of 72 3 10 11 12 Pay attention to number of washers ...
Страница 57: ...page 57 of 72 13 14 15 1 2 Pay attention to number of washers ...
Страница 58: ...page 58 of 72 Replacing the valves and assembling the pump heads 16 17 18 4x ...
Страница 59: ...page 59 of 72 19 21 20 4 6x 4 4 ft lbf 6 Nm Screw in diagonally at first slightly then tighten ...
Страница 60: ...page 60 of 72 22 23 16 15 24 1 2 Maintain all pump heads in the same way 15 16 mm size 2 2x ...