Assembly and connection
IMPORTANT!
4.3
4.3 Vacuum connection
Vacuum connection
WARNING
Risk of bursting
>
Prevent uncontrolled overpressure, e. g., when
connecting to a locked or blocked tubing system.
>
Maximum admissable pressure at vacuum sensor:
1,5 bar/1126 Torr (abs.).
>
Pollution and damages, especially at the flange, could affect
the measurement.
Connection options
Connection via PTFE hose DN 8/10
or
Connection via hose nozzle DN 6/10
or
Connection via small flange KF DN16
Connection via hose nozzle
Required connection material:
Hose nozzle DN 6/10 mm, kurled
nut M14x1, locking ring; optionally: Vacuum hose and compatible
hose clamp (tool: fork wrench size 17).
1.
If installed, unscrew the small flange from the vacuum con‑
nection of the gauge.
2.
Connect hose nozzle
(a)
, kurled nut
(b)
and locking ring
(c)
as shown in figure.
(c)
(a) (b)