background image

 

 

 

   

 
 

 

Page 8 of 13 

 

[email protected] 

www.vacgen.com 

  +44 (0) 1424 836 333

 

 

5.1 Construction

 

 
The  construction  is  based  on  the  standard  double  bellows  universal  type  wobble  stick  to  which  an 
additional co-axial bellows has been added. The movement of this third bellows is used to operate the 
jaws of the pincer grip, the motion being transmitted along the axis of the wobble stick by means of a 
push  rod.  The  jaws  of  the  pincer  grip  are  closed  by  the  action  of  two  springs.  One  of  these  acts 
directly  on  the  jaws;  the  other  retracts  the  push  rod  and  resists  the  effect  of  atmospheric  pressure 
acting inside the bellows. 

 

The  opening  of  the jaws  can  be  adjusted  to  some  extent,  and  the  whole  pincer  grip  can  be  rotated 
about the main axis. It is fitted with detents at 90

o

 intervals thus enabling the plane of the specimen to 

be varied. The rotation is accomplished by bringing the projecting upper part of the jaw assembly into 
contact with a suitably positioned fixed finger or post; the application of a force sufficient to overcome 
the spring-loaded detent will allow the jaw assembly to rotate to the desired position. 
 

5.2 Preparation

 

 
The  only  preparatory  work  required  is  to  ensure  that  the  specimen  is  firmly  held  by  the  jaws  of  the 
pincer grip and that a suitable, but not excessive, degree of opening is achieved. The opening of the 
jaws is controlled by the adjusting screw in the moving jaw. The adjustment should be such that the 
specimen is firmly gripped and registered in the segmental recess in the jaws. Too large an opening, 
especially with a thin specimen, can result in the specimen being able to enter too deeply between the 
jaws which may result in a fragile specimen being damaged. 
 
If  it  is  desired  to  use  the  rotational  facility  of  the  pincer,  the  suitability  of  the  setting  of  the  spring 
pressure on the detent should be checked by a trial rotation. The setting should be such as to give a 
positive  location  whilst  at  the  same  time  allow  the  jaws  to  be  rotated  without  the  application  of 
excessive force. Adjustment of the detent spring pressure is achieved by means of the set screw and 
locknut located at the top of the jaw assembly. 
 
It is also necessary to correctly orient the cassette capstan post (stub) so that the specimen is aligned 
in a plane parallel to the axis of the wobble stick. This must of course be done prior to evacuation of 
the system, and after the pincer grip wobble stick has been installed. If the plane of the specimen is 
not  properly  aligned,  the  removal  and  replacement  of  the  specimens  in  the  cassette  will  be  made 
more difficult and specimens may be dropped. It is recommended that provision is made to catch any 
specimens  that  may  be  dropped,  especially  when  retrieval  is  difficult,  or  problems  could  be  caused 
(e.g. if specimens were dropped onto a gate valve, or into a pump). 
 

Содержание WS Series

Страница 1: ...he part of VACGEN No part of this document may be copied or distributed transmitted transcribed stored in a retrieval system or translated into any human or computer language in any form or by any mea...

Страница 2: ...stallation and Use of Conventional Universal Wobble Sticks 2 4 0 Construction 2 4 1 Transfer Fork 3 4 2 Retractor Cradle 3 5 0 Pincer Grip Wobble Sticks 3 5 1 Construction 3 5 2 Preperation 4 5 3 Inst...

Страница 3: ...clearance or safety certificate in a form satisfactory to VACGEN has been completed and returned by the customer REPAIR The following additional terms and conditions apply in the event that the custom...

Страница 4: ...a much wider range of angular and linear travel WS50 120 182 240 These are often used with a particular design of sample transfer arrangement to lift and carry samples within the vacuum system Access...

Страница 5: ...ensure that the bellows are not damaged in any way by driving them into the port tabulation or chamber wall when operating the device It is equally important not to apply any twisting forces to the b...

Страница 6: ...signed to lock the wobble stick motion when the device is not being used The parking plate locks over the wobble stick shaft and prevents the wobble stick being sucked into the vacuum chamber under th...

Страница 7: ...able as an assembly from the wobble stick thus offering the possibility of alternate grips or end fittings being used Figure 5 The WS75 130 wobble sticks A cassette capable of holding up to 11 un moun...

Страница 8: ...d The opening of the jaws is controlled by the adjusting screw in the moving jaw The adjustment should be such that the specimen is firmly gripped and registered in the segmental recess in the jaws To...

Страница 9: ...tan which engages with a spring loaded detent incorporated in the cassette The cassette also has 6 slots on the periphery into which the nose of the pincer grip jaws can be engaged when it is desired...

Страница 10: ...mm AF spanner on the two flats at the end of the bellows assembly insert a 2 5 mm diameter rod through the transverse hole near the end of the shaft and remove it by unscrewing right hand thread 4 Uni...

Страница 11: ...h interlocks with the pincer grip jaws to ensure that the cassette cannot be accidentally detached The cassette can be lifted off the grooved boss which is secured to its fixing plate by means of a si...

Страница 12: ...Transfer fork 76mm long ZWS20C Retractor cradle for WS20 wobble stick ZWS75C Retractor cradle for WS75 wobble stick ZWS120C Retractor cradle for WS120 wobble stick ZWS182C Retractor cradle for WS182...

Страница 13: ...pment YES NO Toxic YES NO Corrosive YES NO Explosive YES NO Biological Hazard YES NO Radioactive YES NO Other Harmful Substances Equipment and Components that have been contaminated WILL NOT be accept...

Отзывы: